6 Troubleshooting
889 IC Sample Center
■■■■■■■■
41
6.1.3 Syringe unit
Problem
Cause
Remedy
Error 324
The syringe valve has not
reached the required port.
Call the Metrohm Service.
Error 330
The Home sensor of the syringe
has not been activated.
1. Check whether the flow path is kinked
or blocked in on one or more places.
2. If possible, free the flow path of block-
ages.
If the error occurs again, please call the
Metrohm Service.
Error 331
The Home sensor of the syringe
has unexpectedly been activa-
ted.
Call the Metrohm Service.
Error 332
The required load volume of
the syringe is too high.
The permitted load volume is limited by
the maximum syringe volume. The load
volume currently possible depends addi-
tionally on the current position of the
syringe plunger.
■
Ensure that the required load volume
of the syringe is lower or the same as
the maximum syringe volume.
■
Check whether the required load vol-
ume can be reached from the current
position of the syringe piston.
■
Before loading the syringe eject
enough volume first or enter a lower
load volume.
Error 333
The required unload volume of
the syringe is too high.
The permitted unload volume is limited by
the maximum syringe volume. The unload
volume currently possible depends addi-
tionally on the current position of the
syringe plunger.
■
Ensure that the required unload vol-
ume of the syringe is lower or the
same as the maximum syringe vol-
ume.
■
Check whether the required unload
volume is possible assuming the cur-
rent position of the syringe piston.