5.3 Maintenance and service
■■■■■■■■■■■■■■■■■■■■■■
36
■■■■■■■■
889 IC Sample Center
Note
If you are using the sample racks with 12 or 48 vials, then you must
adjust the setting of the needle height to > 2 mm. This will prevent
the needle from touching the bottom of the vial.
9
Perform a rinsing procedure in order to clean the new needle. On the
General
tab, under
Wash
, click on
Start
. To interrupt the rinsing pro-
cedure, click on
Stop
.
5.3.4
Replacing the fuses
The fuses installed in the 889 IC Sample Center are of the type
2 × 2.5 A
Warning
The device must be disconnected from the electricity source before the
fuses are replaced.
Make sure that the fuses to be installed correspond to the correct type and
the correct power load.
The fuses are located in the fuse holder at the rear of the device.
Note
Contact a service technician in the event of repeated problems with the
fuses.