Metrix PX 110 Manual Download Page 5

22

Wattmeters

1.

GENERAL INSTRUCTIONS

Thank you for purchasing this 

PX 110 

or 

PX 120 wattmeter

.

This instrument complies with the NF EN 61010 safety standard concerning electronic
measurement instruments. For your own safety and to prevent damage to the instrument, it
is essential to respect the instructions given in this manual. The contents of this manual must
not be reproduced in any form without our consent.

1.1

Precautions and safety measures

1.1.1

Before using

* This instrument can be used for testing category-III installations, with voltages which

never exceed 600 V (AC or DC) in relation to the earth.

* Definition of installation categories (see publication IEC 664-1):

CAT I:

Category I circuits are circuits protected by low level transient over-voltage
limiters
Example: protected electronic circuits

CAT 

II

 : CAT 

II

 circuits are household appliance or similar power circuits, which may

carry medium-level transient over-voltages.
Example: household appliance and portable tool power supplies

CAT III : CAT III circuits are high-power appliance power circuits, which may carry

high-level transient over-voltages.
Example: industrial machinery or instrument power supplies

CAT IV : CAT IV circuits are circuits which can carry very substantial transient over-

voltages.
Example: power feeder

* To use this instrument, the user must comply with the usual safety rules in order to:

- ensure protection against electric currents,
- protect the instrument against incorrect operations.

* For your safety, only use the leads and the fuse delivered with the instrument: they

comply with the norm EN 61010-1. Before using, always check that they are in
perfect working order.

1.1.2

When using the instrument

* Never exceed the protection limit values indicated in the specifications for each type

of measurement.

* When the instrument is connected to the measurement circuits, do not touch any

unused terminals.

* When repairing TVs or measuring on power switching circuits, there may be high-

amplitude voltage surges on the measuring points which can damage the instrument.
These surges can be attenuated by using a filtering probe.

Chapter I

北京海洋兴业科技股份有限公司(证券代码:839145)

电话:010-62176775

Summary of Contents for PX 110

Page 1: ...PX 110 PX 120 839145 010 62176775...

Page 2: ...LD SMOOTH INRUSH PF kW kVAR kVA O PX 110 POWER METER ON OFF DISPLAY INRUSH SMOOTH HOLD Se reporter au 2 Description See 2 Description Siehe 2 Beschreibung Vedi 2 Descrizione Remitirse al 2 Descripci n...

Page 3: ...rter au 2 Description See 2 Description Siehe 2 Beschreibung Vedi 2 Descrizione Remitirse al 2 Descripci n HOLD T3FE SMOOTH INRUSH PF kW kVAR kVA O PX 120 POWER METER T3FE ON OFF DISPLAY INRUSH SMOOTH...

Page 4: ...RACTERISTICS 26 3 1 Reference conditions 26 3 2 Voltage measurements V 27 3 3 Current measurements A 27 3 4 Power measurements 28 3 5 Communication option 32 3 6 Mains power supply option 32 4 GENERAL...

Page 5: ...nd portable tool power supplies CAT III CAT III circuits are high power appliance power circuits which may carry high level transient over voltages Example industrial machinery or instrument power sup...

Page 6: ...anyone to use it until it has been checked It is advisable to remove the batteries if the instrument is not going to be used for a long time 1 2 Protective systems This instrument is equipped with sys...

Page 7: ...T The PX 110 and PX 120 wattmeters can be used to measure AC and DC power levels voltages and currents They have been designed specifically for general and technical colleges installers and maintenanc...

Page 8: ...bling the instrument to be set up in a tilted position The mains power supply option which is connected in place of the battery hatch 2 2 Display The liquid crystal display allows the display of numer...

Page 9: ...on components by current overload The field of application is limited to industrial frequencies 40 Hz 70 Hz 3 IMPLEMENTATION FUNCTIONAL CHARACTERISTICS The instrument is activated and shut down by pre...

Page 10: ...2 counts Accuracy for DC 1 of reading 3 counts Resolution 0 1 V Display stability 5 cts in normal mode 2 cts in SMOOTH mode Repeatability of the measurement 0 2 typical Display response time 500 ms i...

Page 11: ...SMOOTH mode 5 cts in normal mode 2 cts in SMOOTH mode Response time 500 ms in normal mode of the digital display 3 s in SMOOTH mode 1 Above 10 A 3 the display indicates OL 3 3 3 Characteristics in INR...

Page 12: ...right complying with the polarities indicated red lead on the terminal and black lead on the COM terminal Set up as indicated in the connection diagrams in Fig 4 4 1 single phase or Fig 4 4 2 balanced...

Page 13: ...VA Characteristics Display range 1000 VA 6 kVA Measuring range 10 0 VA 999 9 VA 1000 VA 5999 kVA Accuracy at 50 Hz 1 5 of reading 2 cts 1 of reading 2 cts Resolution 0 1 VA 1 VA 3 4 4 Reactive power...

Page 14: ...e polarities indicated red lead on the terminal and black lead on the COM terminal Connect the measurement leads to the instrument s voltage measurement terminals on the right complying with the polar...

Page 15: ...d with a mains power supply adapter EN 61010 1 600V Cat III 6 9 VA as an option The battery hatch can be removed and replaced with an adapter connected to the AC mains network Connection The adapter h...

Page 16: ...326 1 Electrostatic discharge IEC 61000 4 2 4 kV contact 8 kV air aptitude criterion A industrial site Electromagnetic field IEC 61000 4 3 10 V m aptitude criterion A industrial site Bursts IEC 61000...

Page 17: ...wer 0 7 Current 0 04 Battery voltage 6 5 to 10 V Voltage 0 04 Power 0 1 Common mode Current 0 2 for U 230V 50 Hz Note The term Power covers the active power P and its associated values Q S and PF 5 AC...

Page 18: ...ine batteries taking care to comply with the polarities indicated on the printed circuit Reinstall the hatch and lock the quarter turn screw 6 2 Replacing the fuse Disconnect the leads from the instru...

Reviews: