MI 3200 TeraOhm 10 kV
Measurements
27
-
The result can optionally be saved by pressing the
MEM
key twice, see the chapter
6.1. Store, Recall and Clear Operation.
Legend of displayed symbols:
DIAGNOSTIC TEST
Name of selected function
Fil0 (Fil1, Fil2, Fil3)
Filter type enabled, see the chapter 5.3.
Configuration
5000V
Set test voltage – step 25 V
U=5295
Actual test voltage – measured value
I=55.6nA
Actual test current – measured value
10.5G
Insulation Resistance –
result
C=2.1nf
Capacitance of measured object
Tr:00min 15s
Set timer value
Bar
Analogue representation of Riso result
R15sec=10.6G
Resistance value measured after set time 1
R01min=10.5G
Resistance value measured after set time 2
R10min=10.5G
Resistance value measured after set time 3
DAR=1.67
DAR as ratio of R1min / R15s
PI=1.21
PI as ratio of R03/R02
DD=__
DD result
Notes:
-
A high-voltage warning symbol appears on the display during the measurement to
warn the operator of a potentially dangerous test voltage.
-
The value of the capacitance is measured during the final discharge of the test
object.
-
If enabled, the instrument measures Dielectric Discharge (DD) when the
capacitance is in the range of 5 nF to 50
F.
Setting up the parameters
of the Diagnostic Test:
-
Press
the
SELECT
key, (the Set-up menu appears
on display, see the
Fig. 17
).
-
Select the parameter) to be set using the
and
keys;
-
Adjust
the
parameter using the
and
keys.
-
Complete the set-up adjustments by pressing either
the
ESC
key or
START
key (to run the measurement
directly). The settings displayed last are stored.
Fig. 17. Set-up menu in
Diagnostic Test