Maintenance
12
4 Maintenance
4.1 Replacing fuses
•
F1
M 0.315 A / 250 V, 20
×
5 mm
This fuse protects internal circuitry of low-value resistance function if test probes are
connected to the mains supply voltage by mistake.
•
F2,
F3
F 4 A / 500 V, 32
×
6.3 mm
General input protection fuses of test terminals L/L1 and N/L2.
Warnings:
•
Disconnect any measuring accessory and power off the instrument before
opening battery/fuse compartment cover, hazardous voltage inside!
•
Replace blown fuse with original type only, otherwise the instrument may be
damaged and/or operator’s safety impaired!