MI 3125 InstaltestCOMBO
Insulation resistance
29
Insulation resistance measuring procedure
Select the
INS
function using the function selector switch.
Set the required
test voltage
.
Enable and set
limit
value (optional).
Disconnect
tested installation from mains supply (and discharge insulation as
required).
Connect
test cable to the instrument and to the item to be tested (see figure
5.10).
Press the
TEST
key to perform the measurement (double click for continuous
measurement
and later press to stop the measurement).
After the measurement is finished wait until tested item is fully discharged.
Figure 5.11: Example of insulation resistance measurement result
Displayed results:
R.........................Insulation resistance
Um......................Test voltage – actual value.