www.portableappliancetesters.com.au/support 

Why is the Unit displaying Wrong Voltage on Screen? 

If you are trying to perform a leakage test or PRCD test and ‘Wrong Voltage' appears, it can often be one of three 
reasons: 
  

1. The DeltaGT BT isn't connected to mains power 

2. The fuses have blown in the unit, these are user changeable and are located in the top back section for the unit 

3. The 240v Power supply from the wall is fluctuating or inconsistent and the machine is unable to get a consistent 
power supply to perform the test. 

 

Why won't the DeltaPat connect to the Android Device or Printer? 

The first thing to do is to power off all 3 parts (The Delta tester, Printer and Android Device) and then power on the 
Delta and Android Device and see if they will connect. If they connect you can then power on the Printer. 
  

If you still don't get any communication with the tester, please give us a call on 1300 656 938 and we can walk you 
through more advanced steps to reconnect. 

 

Why is the Zebra Printer printing blank or faded tags? 

This can happen if the printer battery pack is flat - try switching to mains AC power and do a test print. It's also 
important to ensure that the ribbon is in place and doesn’t need to be replaced 
 

My tester has Frozen and wont power off? 

 
You can easily remove the batteries from the back of the unit, just unscrew the back cover panel under the kickstand 
and remove a couple of batteries, this will shut the unit off then you can put the batteries back in and start the unit 
 
 

The system will Print a tag, but my logo will not appear? 

 
Make sure under on the Delta tester under Setup > Communications that the Printer is set to ZebraBT (or Zebra if 
you are using the serial cable to the printer) and not Printek as this will not print the logo on the tag. 
 
 
 

 

Summary of Contents for DeltaPAT MI 3309 BT

Page 1: ...www portableappliancetesters com au support...

Page 2: ...liancetesters com au support Also Included with the kit AC Power Supply for the Zebra Printer Thermal Ribbon installed in Printer Roll of 500 Tags Heavy Duty Case Zebra Bluetooth Adapter Bluetooth Ini...

Page 3: ...like this You can download the app from here https play google com store apps details id si metrel patlink Once the app has downloaded to your device open it and you will get the main home screen Thi...

Page 4: ...the lid of the case This activation provides the full functionality of the software From the main home screen go to Settings Then Device Activation Select the Key with a Plus in the bottom left corne...

Page 5: ...te The printer only needs to be setup once within the software you don t need to connect to the printer each time the connection is between the Android device and the Metrel Tester THe printer only co...

Page 6: ...so needs to be setup to the correct printer this should be set when purchased but if the unit is wiped or deleted you may need to set this up again On the Delta use the UP DOWN keys until the menu sho...

Page 7: ...6 www portableappliancetesters com au support Select ZebraPRN and then it should return you to the communiations screen with PRN NAME ZebraPRN chosen You can now escape back to the main screen...

Page 8: ...the SHORTCODE function in Australia The 6 buttons on the bottom left are the same 6 buttons located on the front of your Metrel Appliance Tester and allow you to navigate and control the machine The F...

Page 9: ...nd wall adaptor The transformer will glow orange during charging and then green when fully charged This can take upto 6 hours to fully charge from flat To run the printer off mains power you will need...

Page 10: ...ntered press the Mem button to save the details then press Test to select that user That user will then appear on all future tags to change the user just go into Setup then User Data and either create...

Page 11: ...e com port may vary depending on computer It will run through each step of the process automatically in this case we had only 1 test saved in the tester It has successfully read the measurements and d...

Page 12: ...l appear on the left hand side Under the project browser on the left hand you will find the new download You can drill down the menus to find each of your tests On the right hand side you will find mo...

Page 13: ...rs com au support Here you can select what report you want to generate You can also enter any relevant data to the report The report will then generate along the top you can then save as a PDF file to...

Page 14: ...ile Clip or Probe is touching clean metal on the body of the appliance you are testing LN Resistance to High or Low Appears on Screen This normally appears when either there is no device connected or...

Page 15: ...can then power on the Printer If you still don t get any communication with the tester please give us a call on 1300 656 938 and we can walk you through more advanced steps to reconnect Why is the Ze...

Page 16: ...software downloads Just simply follow the link below You can also call email or visit our showroom and our Technical Support Staff will be on hand to help with any issue you may have If there is anyth...