METREL A 1513 Instruction Manual Download Page 9

A 1513 Discharge Link 

 

       

Maintenance

 

 

 

 

4  Maintenance 

 
Unauthorized persons are not allowed to open the A 1513 Discharge Link. There are no 
user replaceable components inside the A 1513 Discharge Link. 
 
 

4.1  Cleaning 

 

No special maintenance is required for the housing. To clean the surface of the A 1513 
Discharge Link adapter, use a soft cloth slightly moistened with soapy water or alcohol. 
Then leave the A 1513

 

to dry totally before use. 

 

Warnings:

 

 

 

Do not use liquids based on petrol or hydrocarbons! 

 

Do not spill cleaning liquid over the A 1513! 

 
 

4.2  Service 

 

For repairs under warranty, or at any other time, please contact your distributor. 

Summary of Contents for A 1513

Page 1: ...METREL test and measurement accessories Discharge Link Adapter A 1513 Instruction manual Version 1 1 1 Code no 20 752 541...

Page 2: ...rel si e mail metrel metrel si Mark on your equipment certifies that this equipment meets the requirements of the all applicable EU European Union regulations 2016 METREL No part of this publication m...

Page 3: ...ND OPERATIONAL CONSIDERATIONS 2 2 1 WARNINGS AND NOTE 2 2 2 MARKINGS ON THE A 1513 3 2 3 STANDARD APPLIED 3 3 WORKING WITH A 1513 4 3 1 DESCRIPTION 4 3 2 A 1513 OPERATION 4 3 3 EXAMPLE OF DISCHARGING...

Page 4: ...is unit contains a high ohm discharge resistor 10 kV enclosed in double insulation plastic package A 1513 Discharge Link is special accessory intended for use only with METREL Insulation testers like...

Page 5: ...part of Insulation measurement with Metrel instruments Do not use the A 1513 for primary discharge discharge of tested objects before measurements Never use the A 1513 with voltages higher than rated...

Page 6: ...rd applied The A 1513 Discharge Link Adapter is manufactured and tested according to the following standard listed below Safety LVD EN 61010 1 Safety requirements for electrical equipment for measurem...

Page 7: ...ration To use the A 1513 Discharge link follow these instructions Manual discharge of capacitive load Guarde HV cabel A 1513 Figure 3 2 Manual discharge example Step 1 After the Insulation measurement...

Page 8: ...R s t DUT U 0 0 0 DUT Charged to At time t the voltage on DUT is Voltage U0 Capacity C0 t 0 2 s t 0 5 s t 1 0 s t 2 0 s t 5 0 s Example 1 10 kV 1 F 1 3 kV 67 3 V 0 45 V 0 0 V 0 0 V Example 2 5 kV 10...

Page 9: ...ecial maintenance is required for the housing To clean the surface of the A 1513 Discharge Link adapter use a soft cloth slightly moistened with soapy water or alcohol Then leave the A 1513 to dry tot...

Page 10: ...mensions W x H x L 200 mm x 120 mm x 90 mm Weight 1 44 kg Protection classification reinforced insulation Over voltage category 0 no overvoltages Pollution degree 2 Degree of protection IP 54 Referenc...