![METREL A 1322 Instruction Manual Download Page 27](http://html1.mh-extra.com/html/metrel/a-1322/a-1322_instruction-manual_1779010027.webp)
A 1322 / A 1422
Measurements with A 1422
27
6.5 Welding circuit leakage current
Refer to master instrument Instruction manual, chapter
Single test
measurements
, subchapter
Welding circuit leakage
– Ileak (W-PE)
for detailed
information.
Procedure:
Item
Comment
Prepare test system.
See
chapter 4.3 Working
Connect power supply plug of tested welding
machine into proper test socket.
Connect welding terminals to W1 and W2 terminals
on A 1422.
Enable L-N change parameter on the master
instrument (for single-phase welding machines).
Optional
Start measurement.
Wait until measurement is finished or stop the
measurement.
Check result(s).
Store result(s).
Optional
Figure 6.5:
Welding circuit leakage current measurement
Warning:
During the test, DUT is powered as for normal operation. Pay attention on
possible electric shock!
Note:
Consider any displayed warning before starting measurement!