A 1322 / A 1422
Operation
12
4.3 Working with the adapter
Master instrument automatically recognize properly connected 3P- Active GT / Machine
(Plus) adapter A 1322 / A 1422.
To establish connection, following procedure shall be applied:
Item
Comment
Connect A 1322 / A 1422 and the
master instrument via RS232 interface.
See
chapter 4.1 Connections
Connect the master instrument test
socket and the A 1322 / A 1422 PAT test
socket.
See
chapter 4.1 Connections
Connect the master instrument mains
cord to the output power supply socket
on A 1322 / A 1422.
See
chapter 4.1 Connections
Connect A 1322 / A 1422 to mains
supply.
Turn on fuse/circuit breaker on A 1322 /
A 1422.
The adapter beeps.
Turn on the master instrument.
Select the test function.
See Instruction manual of the master
instrument.
Measuring system is now ready for
testing.
appears on the master
instrument screen for indication of
connected A 1322 / A 1422.
Connect tested equipment to the proper
test socket/plug
See c
hapters: 5 Measurements / 6
Measurements with A 1422
(according to IEC / EN 60974-4)
for
proper connection of appliances /
welding machines to a measuring
system.
Press Start on the master instrument to
start the test sequence.
See Instruction manual of the master
instrument.