USING APOGEE OXYGEN SENSORS WITH ZENTRA SYSTEM
8. In the command box, type
?XAMBR!
9. Select
Send
.
10. Select
Scan
and ensure that the oxygen concentration now reads very close to 20.95%.
CALIBRATION WITH ZL6 AND ZENTRA UTILITY MOBILE
1. Connect the Apogee oxygen sensor cable to an available port on the ZL6.
2. Press the
TEST button on the ZL6 to advertise Bluetooth.
3. Open ZENTRA Utility Mobile on an iOS or Android device.
4. Connect to the ZL6 Bluetooth.
5. Check that the SO-411 is communicating correctly with the ZL6 (an inaccurate oxygen
concentration and a temperature should appear).
6. Go to
More…
7. Go to
Device & Sensor Tools.
8. Select
Apogee oxygen sensor calibration
.
9. Go back to the device main screen and ensure that the oxygen concentration now reads
very close to 20.95%.
If Apogee oxygen sensor calibration is not listed under Device & Sensor Tools, the calibration
can be done using a terminal window.
1. Connect the Apogee oxygen sensor cable to an available port on the ZL6. Note the port
as this will be used as
[Y]
for
2. Press the
TEST button on the ZL6 to advertise Bluetooth.
3. Open ZENTRA Utility Mobile on an iOS or Android device.
4. Connect to the ZL6 Bluetooth.
5. Check that the SO-411 is communicating correctly with the ZL6 (an inaccurate oxygen
concentration and a temperature should appear).
6. Go to
More…
7. Go to
Terminal
.
8. In the Send a Command box, type
sdicmd [Y] ?XAMBR!
, where
[Y]
is the port number that
the oxygen sensor is plugged into.
For example, if the oxygen sensor is plugged into port 2, type
sdicmd 2 ?XAMBR!
9. Select
Send
.
10. Go back to the device main screen and ensure that the oxygen concentration now reads
very close to 20.95%.