INSTALLATION
CALIBRATION WITH ZSC AND ZENTRA UTILITY MOBILE
1. Connect the Apogee oxygen sensor cable to the ZSC device sensor port.
2. Open ZENTRA Utility Mobile on an iOS
®
or Android
TM
device.
3. Connect to the ZSC Bluetooth device.
4. Check that SO-411 data is communicating correctly with the ZSC (an inaccurate oxygen
concentration and a temperature should appear).
5. Go to
More…
6. Go to
Device & Sensor Tools.
7. Select
Apogee oxygen sensor calibration
.
8. Go back to the device main screen and ensure that the oxygen concentration now reads
very close to 20.95%.
If Apogee oxygen sensor calibration is not listed under Device & Sensor Tools, the calibration
can be done using a terminal window.
1. Connect the Apogee oxygen sensor cable to the ZSC device sensor port.
2. Open ZENTRA Utility Mobile on an iOS or Android device.
3. Connect to the ZSC Bluetooth device.
4. Check that the SO-411 is communicating correctly with the ZSC (an inaccurate oxygen
concentration and a temperature should appear).
5. Go to
More…
6. Go to
Terminal
.
7. In the Send a Command box, type
sdicmd ?XAMBR!
8. Select
Send
.
9. Go back to the device main screen and ensure that the oxygen concentration now reads
very close to 20.95%.
CALIBRATION WITH ZL6 AND ZENTRA UTILITY
1. Connect the Apogee oxygen sensor cable to an available port on the ZL6.
2. Open ZENTRA Utility on a computer.
3. Plug the micro-USB cable connector to the ZL6 and the USB cable connector to the
computer.
4. Click the Connect button on ZENTRA Utility to connect to the ZL6.
5. Scan ports to check that the SO-411 is communicating correctly with the ZL6 (an
inaccurate oxygen concentration and a temperature should appear).
6. Go to
Actions
>
Digital Sensor Terminal
.
7. Select the ZL6 port connected to the oxygen sensor cable.