Summary of Contents for Apogee SO-411

Page 1: ...APOGEE OXYGEN SENSOR...

Page 2: ...Introduction 2 2 Installation 2 2 1 Connect to METER ZENTRA Series Logger 3 2 2 Sensor Calibration 4 2 3 Heater Power Supply 7 3 Data Interpretation 7 4 Troubleshooting 7 5 Customer Support 8 Index 1...

Page 3: ...ap gee INSTRUMENTS Made in USA...

Page 4: ...Apogee SO 411 Oxygen Sensor preconfigured by METER included in order Apogee AO 001 Diffusion Head included in order Shovel or Auger if application requires sensor burial METER ZENTRA series data logg...

Page 5: ...er and secure cables between the sensors and the data logger to the mounting post in one or more places to ensure cable weight does not pull the plug free from its port Connect to Data Logger Plug the...

Page 6: ...measurement interval 5 Use ZENTRA Utility to configure communication settings for data transfer to ZENTRA Cloud Sensor data can be downloaded from METER data loggers using either ZENTRA Utility or ZEN...

Page 7: ...A Utility Mobile on an iOS or Android device 3 Connect to the ZSC Bluetooth device 4 Check that the SO 411 is communicating correctly with the ZSC an inaccurate oxygen concentration and a temperature...

Page 8: ...ygen concentration now reads very close to 20 95 If Apogee oxygen sensor calibration is not listed under Device Sensor Tools the calibration can be done using a terminal window 1 Connect the Apogee ox...

Page 9: ...be a simple wall wart If mains power is unavailable a 12 V battery is appropriate and a solar panel and charge controller can be used to maintain battery charge Under most conditions a small battery 7...

Page 10: ...ted periodically to maintain best accuracy See Section 2 2 for directions on calibrating the sensor Refer to the SO 411 User Manual apogeeinstruments com so 411 sdi 12 soil response thermistor referen...

Page 11: ...by email please include the following information Name Address Phone Email address Instrument serial number Description of the problem NOTE For products purchased through a distributor please contact...

Page 12: ...ies data loggers I installation 2 3 above ground 3 buried 3 preparation 2 tools needed 2 internal heater 2 3 7 O oxygen concentration 2 4 oxygen sensor 2 S stereo plug connector 3 T troubleshooting 7...

Page 13: ...ins Court Pullman WA 99163 T 1 509 332 2756 F 1 509 332 5158 E info metergroup com W metergroup com METER Group AG Mettlacher Stra e 8 81379 M nchen T 49 89 1266520 F 49 89 12665220 E info europe mete...