TROUBLESHOOTING 

 
PROBLEM 

POSSIBLE CAUSE 

CORRECTION 

LCD does not 
light 

 

No power input 

 
 

 

Check power cord 
connections 

 

Turn on power 

 

System will not 
actuate 

 

Foot switch not 
plugged in or 
improperly plugged 
in 
 

 

Defective foot switch 

 

 

Broken wire or loose 
connection inside 
unit 

 

 

 

Defective solenoid 

 

Defective PC board 

 

Check foot switch 
connection 

 
 

 

Foot switch needs to be 
repaired or replaced 

 

Unplug power cord and 
disconnect air supply.  
Remove cover and check 
for broken wires or loose 
connections 

 

Replace solenoid 

 

Replace PC board 

System will not 
pressurize 

 

Insufficient air 
pressure 

 

Air hoses not 
plugged in 

 

Regulator defective 

 

Increase air supply 
pressure 

 

 

Check connection 

 

Replace regulator 

System will not 
pull vacuum 

 

Vacuum setting is 
too low 

 

Defective solenoid 

 

Defective vacuum 
venturi 

 

Defective vacuum 
needle valve 

 

Increase vacuum setting 

 

 

Replace solenoid valve 

 

Replace venture 

 

 

Replace needle valve 

Inconsistent 
dispensing 

 

Air bubbles in 
adhesive 

 

Dispense time is too 
low 

 

Dispense needle 
started to clog 

 

Reduce vacuum setting 

 

 

Increase dispensing time 

 

 

Replace needle 

 

 
 
 
 
 
 
 
 
 
 
 

 

8

 

8

 

MAINTENANCE  

 

The dispenser is designed and built to be relatively maintenance free.  To assure 
trouble free operation, the following recommendations should be followed: 

1.

 

Make certain air supply is clean and dry 

2.

 

Avoid turning barrels upside down or laying barrel so that material may 
run through air line to internal components 

3.

 

Avoid connecting the unit to excessive moisture or solvent saturation. 

4.

 

Avoid connecting air supply exceeding 100 psi (6.9 bar) for the DX-350 
and 15 psi (1.03 bar) for the DX-355 

5.

 

Use only Amyl Alcohol to clean outside surface of the main housing. 

6.

 

Use only soft cloth to clean the LCD 

 

9

 

SPARE PARTS  AND ACCESSORIES 

 

For DX-3xx/2xx spare parts and accessories information, please visit the DX-
3xx/2xx User Guide webpage at 

http://www.metcal.com

 

 

 

 

 

10

 

LIMITED WARRANTY 

 
Metcal warrants  this  product  to  the  original  purchaser  for  a  period  of  one  (1)  year 
from  date  of  purchase  to  be  free  from  material  and  workmanship  defects  but  not 
normal  wear-and-tear,  abuse  and  faulty  installation.   Defective  product  or 
subassembly  and  components  under  warranty  will  be  repaired  or  replaced 
(at Metcal's option) free of charge. Customer with defective product under warranty 
must contact the nearest Metcal office or distributor to secure a return authorization 
prior to shipping the product to the assigned Metcal authorized service center.  For 
nearest  Metcal  office  or  distributor  contact  information,  please  visit 
www.metcal.com.   Metcal  reserves  the  right  to  make  engineering  product  changes 
without notice. 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

Summary of Contents for DX-355

Page 1: ...following actions Making changes to equipment that has not been recommended in the User Guide Using incompatible or damaged replacement parts Using unapproved accessories or auxiliary equipment 1 2 Safety Precautions Do not operate this unit in excess of maximum ratings settings Always wear appropriate personal protective clothing or apparel Use only the following power adaptor models PSA15R 240P ...

Page 2: ...0 for temperature up to 31 C 87 8 F Decreasing linearly to 50 relative humility at 40 C 104 F Air Output DX 350 0 100 psi 0 6 9 bar DX 355 0 15 psi 0 1 03 bar Vacuum 406mm 16 of Hg Timer 0 008 60 000 seconds Cycle Mode Timed Interrupt Teach Purge Timing Repeat Tolerance 0 001 Cycle Rate 1 200 cycles min LCD 20 X 4 display segments 4 4 FEATURES Figure 1 0 Items Description 1 Power Button 2 Mode But...

Page 3: ...Depress Foot switch to activate timed dispensed cycle 6 3 Teach Mode Setting In the teach mode the unit will accumulate time as long as the foot switch is depressed This is helpful in determining dispense time required when dispense output is unknown 1 Press the Mode button 2 to select TEACH mode 2 Press and release the reset button 3 until the dispense time 0 000S 3 Press and hold down Foot switc...

Page 4: ...he following recommendations should be followed 1 Make certain air supply is clean and dry 2 Avoid turning barrels upside down or laying barrel so that material may run through air line to internal components 3 Avoid connecting the unit to excessive moisture or solvent saturation 4 Avoid connecting air supply exceeding 100 psi 6 9 bar for the DX 350 and 15 psi 1 03 bar for the DX 355 5 Use only Am...

Page 5: ...DC range and a maximum of 250 mA Note During the dispense cycle pin 3 will be grounded Please make sure the external device your machine that controls the dispenser controller has the same ground as the DX 350 355 controller Pin 7 8 and 9 Available Chasis Ground To initiate a dispense cycle with contact closure When low pressure alarm is triggered Pin 4 is grounded When low pressure alarm is turne...

Reviews: