10. Storage and transport conditions
The META Tower is delivered separately, on a pallet, wrapped in film and
secured with plastic straps.
The META Tower must be protected against moisture, dirt and external
temperatures during transport. The packaged META Tower must be
stored in a dry place so that it is protected against moisture (max.
relative air humidity: 70%). The storage temperature must be between
-10°C and +40°C.
The META Tower must be depressurised when being
transported. Before transporting, the tank‘s compressed
air and the hoses must be bled and condensation drained if
necessary.
11. Installation guidelines
After unpacking the META Tower, it must be inspected carefully in
order to ensure that it is in perfect condition.
The META Tower may only be installed in dry rooms with low dust
levels and sufficient ventilation.
If the ventilation is insufficient, a suitably sized fan or suction
system must be used.
Use in explosive or flammable locations is prohibited.
The permissible room temperature range is b10° C and
+40° C. The relative air humidity must not exceed 70%.
It can be set up in a separate room on the same floor as the treat-
ment stations or one floor below them. For noise sensitivity rea-
sons, we recommend setting up the META Tower in a separate room.
The META Tower must be installed on a solid surface in order to pre-
vent vibrations.
In order to ensure the best possible ventilation for the META Tower,
the gap between limitations on the sides (e.g. cupboards) and the
META Tower must be at least 5 to 10cm.
Install the META Tower so that it is easily accessible for operation
and maintenance purposes.
The META Tower must be installed on a level surface with sufficient
load bearing capacity.