background image

Hygienesystem 

H1

Gerätedokument

Einbau, Betrieb und Wartung

Hygiene system 

H1

Equipment Logbook

Assembly, operation and maintenance

Systeme d’hygiene

H1

Livret d’appareil

Installation, fonctionnement et entretien

Sistema d’igiene 

H1

Verbale d’installazione

Montaggio, funzionamento e manutenzione

H

1

1

EN

Summary of Contents for H1

Page 1: ...d Wartung Hygiene system H1 Equipment Logbook Assembly operation and maintenance Systeme d hygiene H1 Livret d appareil Installation fonctionnement et entretien Sistema d igiene H1 Verbale d installazione Montaggio funzionamento e manutenzione H1 1 EN ...

Page 2: ...on requirements 5 8 Modular assembly 6 9 Functional description 7 10 Cleaning cycles 9 11 Hygiene System H1 maintenance 10 12 Hygiene System H1 control panel 11 13 Electronic circuit board 12 14 Inspections 14 15 Cleaning and disinfection 14 16 Refilling the H1 cartridge 15 17 Disposal 15 2 Explanation of the pictograms i Information Caution General warning sign Refer to instruction manual booklet...

Page 3: ...nance Only original parts may be used for repairs or replacements All the guidelines provided by the manufacturer on the correct use of the Hygiene system H1 must be followed All inspection and service work must be entered into the device docu mentation at the first pages of this manual When requested by an authorized engineer METASYS agrees to make all documents available for the use of technical...

Page 4: ...5 Explanation of the type plate 2 See illustration 2 1 Serial number 2 2 Address of the manufacturer 2 3 Mains supply data 2 4 Protection class II 2 5 Type BF sign 6 Technical data Permissible water pressure 2 4 bar working pressure 2 bar Water connection PU tube nominal width 2 mm external Ø 4 mm Voltage 24 V AC max current consumption 450 mA Frequency 50 60 Hz max power consumption 11 VA Volume ...

Page 5: ...ar at the cannula Voltage to terminal X2 must be 24 V AC Themanufacturer sregulationsregardingtheinstallationofparticular units must be adhered to 3 See illustration 3 1 Suction line 3 2 Suction hose 1 3 3 Suction hose 2 3 4 Vacuum switch 3 5 Connection cable 4 and 6 pole control unit support signal 3 6 Voltage 24 V AC red black start signal for suction system yellow 3 7 Mains water supply PU tube...

Page 6: ...r network the fresh water is taken from a container with free outlet per DVGW 7 Module 3 Filter housing elbow closure pieces suction tubes The filter housing contains a removable non drip rake The elbow closure pieces one per suction tube serve as locking devices for the suction flow and in order to interrupt the supply of disinfectant if vacuum pressure drops They are rotatably hung and form the ...

Page 7: ... 4 Pump chamber 5 Suction flow closure 6 Chemical closure 7 Filter housing 8 Handpiece 9 Suction tube with inner tube 10 Amalgam separator or automatic separator 11 Suction system Lifting the handpiece 8 of a suction hose results in a vacuum causing the respective suction flow sealing membrane 5 to be opened by an associated solenoid valve L R 12a The suction flow seal is designed as a pipe elbow ...

Page 8: ...and through lines in the suction hoses to the handpiece in use The cleaning mixture is then distributed evenly over the inner surface of the suction tube by the suction flow and thus ensures optimal cleaning and disinfection of the whole suction system 16 Adjusting the suction flow rate see illustration The suction flow rate can be adjusted by a regulator above the filter cup A screwdriver or a co...

Page 9: ...on hose Sequence 1 x H1 413 x H2O 1 x H1 per hose This sequence is halted with an interruption to the suction operation and is continued when the suction operation resumes or starts anew 19 Special cleaning The special cleaning program is started by pressing the respective button on the operating interface This program should be used at the end of each treatment day and after particularly intense ...

Page 10: ... operation If the performance of the suction system declines check the filter rake for contamination Changing the H1 cartridge pouch see illustration 27 Please also refer to the H1 instructions on the refill package and on page 11 If the yellow monitor light no 2 empty tank warning on the control panel flashes and a short alarm is audible every time the suction tube is lifted please exchange the H...

Page 11: ... once if this does not work then an electrical fault exists Check electronics 28 2 Control light 2 Empty tank warning yellow light and single beep whenever the suction tubes are lifted Change the H1 cartridge or H1 pouch 28 3 Control light 3 Program in use green light on suction hose in place ready for operation green light on suction hose in use continuous cleaning in progress green light flashin...

Page 12: ...V X2 24V 29 3 Start of suction motor X2 SM 29 4 Suction tube holder 1 2 and 3 X3 ABL 29 5 Sensor chemical tank X4 CH SENS 29 6 Valves V1 V5 X5 VENTILE 6 pole 29 7 Control panel X6 BED TEIL 29 8 Vacuum switch 30 Transition board see illustration V1 H2O V2 GREEN CLEAN H1 V3 Knee valve right V4 Knee valve left V5 Additional valve for third suction hose Electronic circuit board Technicians ...

Page 13: ...ions which are not of type BF a cut off relay that complies with IEC 601 1 section 20 2 insulation B a must be used This cut off relay is switched through the output contact X2 whereupon the operating voltage used must be the same as of the H1 system 31 1 H1 mode setting 31 2 SIC1 fuse 750m A T wired in 32 H1 mode setting see illustration Switch 1 32 1 Standard 1 two suction hoses 32 2 3 suction h...

Page 14: ...tube 33 1 Handpiece small order no 50010047 Large suction tube 33 2 Handpiece large order no 50010048 33 3 Distributor large Order no 50010001 Grease the o rings with vaseline after cleaning 34 Prefilter see illustration 34 1 Filter rake order no 55010001 34 2 Filter cup order no 50010008 35 Lubricate the sealing lip with vaseline and place back properly The entire suction tubes can be disinfected...

Page 15: ...er to the refill pouch Attach the long end of the adapter to the cartridge 40 Refill the cartridge Fill up the cartridge with GREEN CLEAN H1 until the maximum level is reached Remove adapter after use and rinse with water Close the refill pouch Insert plunger with cartridge seal in filling opening Attach the bayonet socket and turn the key clockwise 17 Disposal When the complete device is de insta...

Page 16: ...205420 7 www metasys com info metasys com METASYS makes the difference Visit us at GERMANY 49 8823 938 44 33 info metasys com FRANCE 33 4 37 90 22 15 info metasys fr ITALY 39 045 981 4477 desitalia metasys com 2019 05 H1 55 005 01 Druck und Satzfehler vorbehalten Subject to printing and setting errors ...