16. Maintenance, cleaning and disinfection
29
See illustration
Shortly operate the spittoon bowl after every treatment!
30
See illustration
Suck off some water with each of the suction tubes after every treatment.
31
See illustration
Twice a day, after having sucked off some water, use the disinfectant
for suction systems and amalgam separators GREEN
&
CLEAN M2 (item
no: 60010201) or the cleanser GREEN
&
CLEAN CL (item no: 60010301).
Ideally a disinfection with GREEN
&
CLEAN M2 or a cleaning with
GREEN
&
CLEAN CL should be carried out before longer periods of
downtimes of the dental unit (e.g. lunch break, end of the working
day or holidays).
32
See illustration
The spittoon bowl should also be rinsed with GREEN
&
CLEAN M2 or
GREEN
&
CLEAN CL twice a day..
The operation of the amalgam separator COMPACT Dynamic without
prefilter at the dental unit and/or at the central suction system is not
permitted!
Cleaning the prefilter:
The prefilter must be emptied and cleaned at least once a week. This may
also be carried out daily, depending on workload.
33
See illustration
The amalgam residue from the prefilter box must be collected in the
METASYS
ECO CENTER
and disposed of properly with
ECO TRANSFORM
.
Maintenance, cleaning and disinfection
Practice personnel, technicians