10
b.
Testing
the
Current
Loop
i.
Calibration
The
MS2600E
is
fully
calibrated
when
shipped
from
the
factory.
The
calibration
settings
are
fixed
to
avoid
accidental
change
which
could
result
in
erroneous
data.
No
field
calibration
is
required.
However,
it
is
important
to
test
the
MS2600E
upon
installation,
and
during
periodic
maintenance
inspections,
to
ensure
the
unit
is
operating
properly.
ii.
Testing
loop
output
zero
(4mA)
and
span
(20mA)
The
MS2600E
Probe
Selection
Switches
(page
9)
offer
three
test
settings
that
allow
the
unit
to
be
placed
into
various
diagnostic
modes
as
follows:
1)
4mA
Output
–
Forces
a
constant
4mA
output
on
the
current
loop.
2)
20mA
Output
–
Forces
a
constant
20mA
output
on
the
current
loop.
3)
4
‐
20mA
Sweep
–
Causes
the
output
to
continually
cycle
from
4mA
up
to
20mA.
These
test
modes
can
be
useful
when
troubleshooting
problems
with
the
current
loop
wiring
and
DCS/SCADA
system.
iii.
Testing
the
MS2600E
with
the
Meter
Prover
A
Meter
Prover
is
provided
to
allow
routine
checks
of
the
MS2600E.
The
Meter
Prover
simulates
a
Wire
Loop
type
probe
at
a
fixed
value.
To
test
the
MS2600E
with
the
Meter
Prover:
1)
Disconnect
power.
2)
Disconnect
the
MS2600E
from
the
probe
(or
if
the
MS2600E
is
remote
‐
mounted,
disconnect
the
probe
extension
cable
from
the
probe.)
3)
Connect
the
Meter
Prover
to
the
MS2600E
probe
connector
stem
(or
to
the
probe
extension
cable
if
the
MS2600E
is
remote
‐
mounted.)
4)
Loosen
the
MS2600E
Enclosure
Lock
Screw.
5)
Unthread
and
remove
the
MS2600E
cover.
6)
Change
the
Probe
Selection
Switches
to
the
Wire
Loop
position
(see
page
9.)
7)
Reconnect
power
and
allow
the
instrument
to
measure
for
several
minutes
to
stabilize.
8)
After
several
minutes
observe
the
transmitter
output.
The
output
should
closely
match
the
value
printed
on
the
Meter
Prover
label.
If
the
transmitter
output
matches
the
Meter
Prover
value,
you
may
return
the
Probe
Selection
Switches
to
their
previous
setting
and
reconnect
the
MS2600E
to
the
probe.
If
the
transmitter
output
shows
a
significant
difference
compared
to
the
Meter
Prover
value,
further
troubleshooting
may
be
required.
Refer
to
page
14
for
troubleshooting
or
contact
the
factory
for
further
assistance.
Be
sure
to
reinstall
the
enclosure
cover
and
tighten
the
Lock
Screw
when
putting
the
MS2600E
back
into
service.