D. Probe-Instrument Connection
The MS3500E/3510E can be mounted directly onto an ER probe, or onto an HA700110-5 probe
adaptor in the case of a retrievable probe used in a two-inch, high-pressure, access fitting (see Figure
12, p. 41). However, in many instances it is either inconvenient or undesirable to mount the instrument
directly onto the probe. For these circumstances a 546810, ten-foot, extension cable is provided. The
extension cable has integral connectors with a guiding keyway to ensure correct pin alignment when
attached. The female connector (MS3106E145-6S) mates with the probe, or HA700110-5 probe
adaptor, and the male connector mates with the central terminal on the base of the MS3500E/3510E
unit.
The length of the extension cable has been carefully determined so as to avoid excessive
signal noise. Longer extension cables should only be employed after consultation with Metal
Samples.
The extension cable provided will mate directly with any Metal Samples’ ER probe or those of any
other major manufacturer, or with an HA700110-5 probe adaptor in the case of retrievable probes.