7.2.
Real Time Output
The real time output occurs when the unit finishes a sample. The output format is
either a comma separated value (CSV) or a printer style depending on the Serial Report
mode.
7.3.
Comma Separated Value (CSV)
The CSV output fields are both comma separated and fixed length.
CSV Header (Note 1):
Time,Size1,Count1(M3),Size2,Count2(M3),Size3,Count3(M3),Size4,Count4(M3),Size5,
Count5(M3),Size6,Count6(M3),AT(C),RH(%),Location,Seconds,Fav1Size,Fav2Size,Status
CSV Example Record:
2013-09-30
10:04:05,00.3,08562345,00.5,01867184,00.7,00654892,01.0,00245849,02.0,00055104,05.0,00
023,040,001,010,00.3,00.5,000,*00086
CSV Fields
Field
Parameter
Example Value
Notes
1
Date and Time
2013-09-30 10:04:05
2
Channel 1 Size
0.3
3
Channel 1 Count (TC, /L, CF, M3)
8562345
Note 2
4
Channel 2 Size
0.5
5
Channel 2 Count (TC, /L, CF, M3)
1867184
Note 2
6
Channel 3 Size
0.7
7
Channel 3 Count (TC, /L, CF, M3)
654892
Note 2
8
Channel 4 Size
1.0
9
Channel 4 Count (TC, /L, CF, M3)
245849
Note 2
10
Channel 5 Size
2.0
11
Channel 5 Count (TC, /L, CF, M3)
55104
Note 2
12
Channel 6 Size
5.0
13
Channel 6 Count (TC, /L, CF, M3)
31790
Note 2
14
Temperature (C,F)
23
Note 2 & Note 3
15
RH (%)
40
Note 3
16
Location
1
17
Sample Time (0-9999 seconds)
60
18
Favorite 1 Size
0.3
Note 4
19
Favorite 2 Size
0.5
Note 4
20
Status Bits (see below)
0
Note 5