AEROCET 532 Manual Rev B
Page 29
7.2.2. Mass Format
Mass Data Report
2021-05-10 08:40:14
Serial Number, B12561
Time,PM1(ug/m3),PM2.5(ug/m3),PM4(ug/m3),PM7(ug/m3),PM10(ug/m3),TSP(ug/m3),AT(
C),RH(%),Location,Seconds,Status
2021-05-07 15:39:09,001.6,001.6,001.7,001.8,001.8,001.9,+024.9,030,LOC1
,0060,0000
CSV Fields
Field
Parameter
Example Value
1
Date and Time
2021-05-07 15:39:09
2
PM1 (µg/m3)
001.6
3
PM2.5 (µg/m3)
001.6
4
PM4 (µg/m3)
001.7
5
PM7 (µg/m3)
001.8
5
PM10 (µg/m3)
001.8
6
TSP (µg/m3)
001.9
7
AT (C, F)
+024.9
8
RH (%)
030
9
Location
LOC1
10
Seconds
0060
11
Status
0000
7.2.3. Status
The last entry of the CSV output is the status indicating alarms or errors. Status bits
combinations are possible. For example, 18 = IOP Alarm and Temperature Sensor
Alarm.
Status Bits
Bit
Value
Condition
0
OK (no alarms or errors)
0
1
Not used
1
2
IOP Alarm (Laser)
2
4
Not used
3
8
Not used
4
16
Temperature Sensor Alarm
5
32
Pressure Sensor Alarm