1.3
Safety Instructions
• For safe operation of the device please pay regard to all instructions and warnings in this manual
• Only put the device into operation after it has been installed properly. A competent and authorised
person is required for installation, connection and operation of the device. This person has to read
the manual and follow all instructions. Keep this manual to look up questions that can occur later
on.
• Defective devices must be disconnected from the process! This applies for apparent damages
of the device such as physical damages but also in the case of unclarified malfunctions in the
operation. Separate the device from the process pneumatically (both gas inlet and gas outlet)
and remove the power supply from the device.
• Make sure that the electric installation protection against accidental contact agree to the applica-
ble safety regulations. The protective earth connection must be made before all other connec-
tions. Any interruption in the protective earth can cause danger.
• Pay regard to the local regulations and circumstances regarding electric installations.
• Repairs may only be done by Messkonzept.
1.4
Safety instructions for use in explosion hazard areas Zone 2
Conformity with DIN EN 60079-15 "Explosive atmospheres - Part 15: Equipment protection by type of
protection "n" : 2011-02 and 2020-03" was proven with report no. 2.12BR190104AUG2V06 "Bericht
zur Eignung des Standardgehäuses der Messgeräte FTC3xx und FTC4xx für Zone 2, Schutzart „nR“,
schwadensicheres Betriebsmittel"
. If flammable gas is fed into the analyzer at overpressure, a leak in
the gas path could result in an ignitable mixture in the analyzer housing. To prevent the formation of an
explosive atmosphere in the device itself even in the event of a leakage, the housing is internally filled
with glass beads which, due to their filling density, leave only a small free volume of about 0.15l in the
housing. Moreover, due to the small size of the glass beads, this free volume is fragmented into such
small sub-volumes that a flame cannot spread.
Therefore the device may only be opened by the
manufacturer.
1.5
Intended Use
Only gases that are non-corrosive and free of condensate, dust, aerosol or oil mist may be lead in the
FTC-series gas analyzer. Flammable gases require appropriate protective measures. Explosive gases
may not be lead into the FTC. The Instrument may not be used in hazardous areas. Please contact
[email protected] for detailed information and solutions.
1
Report on the suitability of the standard housing of the FTC3xx and FTC4xx measuring instruments for Zone 2, protection
type "nR", steam-proof equipment
File name:
FTC320Ex Operating Manual_1.09KD201009MPO3V02.pdf