The T90-time is the time, in which a sudden change of the measurand (e.g. the gas
concentration) reaches 90% of it’s final value.
6.4
Output Setup
0.30
[%]
H2/N2
Current Loop
Analog Out 1
Analog Out 2
Relay Setup
Diagnosis
Instr. Setup
Output Setup
Expert Setup
Figure 6.6: Analog output setup menu
The FTC320 is equipped with three analog outputs. “Current Loop“ is an isolated current output. “Ana-
log Out 1“ and “Analog Out 2“ are not DC-isolated outputs with an output range from 0 to 10 V. In the
”Output Setup” menu the analog outputs can be configured.
As part of a recent firmware upgrade, the settings of the analog outputs and relays have been restruc-
tured. The new menu structure does not yet have a complete user manual. Please contact Messkonzept
with your questions about the analog outputs and relay settings.
6.5
Expert Setup
0.30
[%]
H2/N2
Diagnosis
Instr. Setup
Output Setup
Expert Setup
Expert Code
set: 0.00000
new: *******
ESC/
OK
Figure 6.7: Expert Setup menu
The Expert Setup provides a couple of functionalities that should only be used by an advanced user or
expert:
• set parameters
• reset to factory settings
• change the “Operator Code“ and the “Expert Code“
• swap between “Normal Mode“ and “Safety Mode“
• simulate alarms and analog outputs
The settings explained here are for advanced users or experts and should not be entered
by normal operators. It is in the responsibility of experts to set the parameters properly.
The default expert code is “222.000“.