41
activating keys have to be pressed simultaneously with the dispense of the start reagent). The
measurement is stopped automatically when a coagulation reaction starts respectively when a
clot is formed.
According to the parameter and the settings the result can be converted to another result unit.
Please regard the instructions (chapter 4) for pipetting plasma and reagent.
For preparing the next measurement press the activating key below the according measuring
position or reset all measuring positions simultaneously by pressing the F3-key (reset of all
measuring positions with the F3-key is not supported during automatic channel reset(chapter
3.5.9)).
5.4
Measurement of parameters with two reagent components
On the whole measurements of parameters with two reagent components (e.g. aPTT) do not
differ from measurements of parameters with one reagent component. Nevertheless you can
pipette the first reagent into the measuring cuvettes when these are still in the pre-heating
positions above the measuring positions, i.e. before the cuvette is changed over to the
measuring position. Please regard the parameter-related incubations time and then proceed as
described in chapter 5.3.