Operating Instructions - CPC8000
Table 10.2 - Troubleshooting
Type
Problem
Correction Action
1
The system has been switched on and no
measurement(s) have appeared and the
entire area of the screen is white (or dark).
Switch the system of. Wait 5 seconds and
switch the system on.
2
The screen is dark and the corrective action
for #1 above has no efect
Check that the power cable is connected
properly and have authorized technical staf
check that the supply voltage is correct.
3
The screen is dark and the action taken to
solve the problem of #2 above has no efect.
Unplug the unit from the power source (wall
socket), then remove the power cord from
the instrument. The power cord input socket
includes two internal fuses. Check both fuses
for continuity.
Fuse holder
Power cord input socket
WARNING! When needed, use
only 2.5 amp 250V SLO-BLO
5x20 Fuse
4
Malfunction during operation.
Switch the system of. Wait 5 seconds and
switch the system on.
5
Unstable control.
Consult factory
6
The setpoint value is not reached.
Check whether the value of the supply
pressure is the value required, and leak test
the plumbing.
If you need help or assistance, contact Mensor at
Phone: 1.512.396.4200 or 1.800.984.4200 (USA only)
Web site: www.mensor.com
Fax: 512.396.1820
E-mail: [email protected]
Summary of Contents for CPC8000
Page 6: ...High End Pressure Controller CPC8000 6 Operating Instructions CPC8000 NOTES ...
Page 10: ...High End Pressure Controller CPC8000 10 Operating Instructions CPC8000 NOTES ...
Page 24: ...High End Pressure Controller CPC8000 24 Operating Instructions CPC8000 NOTES ...
Page 84: ...High End Pressure Controller CPC8000 84 Operating Instructions CPC8000 Notes ...
Page 112: ...High End Pressure Controller CPC8000 Operating Instructions CPC8000 113 NOTES ...