Operating Instructions - CPC3050
19
High-Speed Pressure Controller
CPC3050
4.2 Base Instrument
Instrument
Instrument version
Standard: desktop case
Option: 19” rack-mounting kit
Dimensions
See technical drawings
Weight
approx. 26.4 lbs (12 kg) with all internal options
Warm-up time
approx. 15 min
Display
Screen
7.0” color LCD with resistive touchscreen
Resolution
4 … 6 digits depending on range and units
Connections
Pressure connections
5 ports with 7/16”- 20 F SAE and 1 port with 10-32 UNF female
Filter elements
The instrument has a 40-micron filters on all pressure ports.
Pressure port adapters
Standard: without
Option:
6 mm tube fitting, 1/4” tube fitting, 1/4” female NPT fittings, 1/8” female NPT fittings or 1/8” female
BSP fittings
Barometer port adaptors
Standard: barb fitting
Option: 6mm tube fitting, 1/4“ tube fitting
Permissible pressure media
Dry, clean air or nitrogen (ISO8573-1:2010 Class 5.5.4 or better)
Wetted parts
Aluminum: 6061-T6 / 2024-T4, Stainless Steel: 303 / 304 / 316 / 316L / 13-8 PH, Brass, Nickel,
Buna-N, FKM/FPM, Tungsten carbide, Cerazur ceramic, PTFE, PEEK. Glass-filled epoxy, RTV,
ceramic, silicone, silicone grease, urethane, 430FR, 300 series CRES.
Overpressure protection
Safety relief valve for regulator 3,600 psi.
Permissible pressure
Supply port
110-120% FS or 0.69 bar (10 psi), whichever is greater. (Max limit of 3,350 psi / 231 bar)
Measure/Control port
max. 105 % FS
Voltage supply
Power supply
AC 100 ... 120 V, 50/60 Hz; AC 220 ... 240 V, 50/60 Hz
Power consumption
max. 150 VA
Permissible ambient conditions
Storage temperature
-20 ... 70 °C (-4 ... 158 °F)
Humidity
5 … 95 % r. h. (relative humidity, non-condensing)
Compensated temperature range
0 … 50 °C (50 ... 122 °F)
Mounting position
horizontal
Control parameter
Control stability
High Speed: < 0.025% FS of the primary range, Industrial: <0.01% FS of the primary range
Control mode
Industrial, High Speed and Custom
Control time
3-4 s (regarding a 25% FS pressure increase above atm. in high speed mode)
Control range
0 ... 100% FS
Minimum control pressure
0.0017 bar (0.025 psi) over exhaust pressure or 0.05% FS whichever is greater
Overshoots
< 1% FS in high speed control mode (typical <0.5% FS in industrial control mode)
Test volume
0 ... 1,000 cc
Communication
Interface
Standard: Ethernet, IEEE-488, USB, RS-232.
Command sets
Mensor, WIKA SCPI, PACE, other options
Update time
approx. 30 ms
Internal program