34
Environmental Conditions
Ingress Protection rating: IP40
Protected against foreign objects larger than 1mm,
ingress of dust is not totally prevented. Device is not
protected against water.
Water resistant rating: Main unit: IPX0 – Not water resistant or protected
Condition
Operating
Non-Operating
(Storage and transportation)
Temperature +5℃ to +40℃*
-30℃ to +80℃
Relative
Humidity
10% to 80%, non-condensing
0% to 90%, non-condensing
Maximum
Altitude
3,048 meters (20,000 feet)
12,000 meters (40,000 feet)
Atmospheric
Pressure
>69 kPa & <132 kPa
>23 kPa & <132 kPa
* Wait for at least 10 minutes before use from the extreme non-operating temperature.
5.2 Protocols and Frequency pairs
MEND IS02PROs provides preloaded protocols and can store up to 999 customer
protocols. Long descriptions for each protocol can impact the number of protocols
the customer can load. Each protocol can be programmed to have one or multiple
frequency pairs. Each frequency pair includes output parameters of frequencies for
two channels, current intensity, time, and all other parameters for one therapy
sub-session. The factory default protocols are listed in Table 2.