54
Sterilisers SFplus/SNplus
8. Sterilisers SF
PLUS
/SN
PLUS
8.1 Intended use
The SF
PLUS
/SN
PLUS
PLUS
PLUS
appliance serves for sterilising medical material through dry heated air at
atmospheric pressure.
8.2 Note in accordance with Medical Devices Directive
The product lifetime as intended by the manufacturer is eight years.
8.3 Guidelines for sterilisation
For hot air sterilisation, there are different guidelines on the temperature and sterilisation time
to choose, as well as on packaging the sterilisation load. The values to be chosen depend on
the type and characteristics of the load to be sterilised and on the type of germs to be neutral-
ised. Before beginning sterilisation, make yourself familiar with the sterilisation method laid
down for your application.
Process parameters for hot air sterilisers are temperature and minimum hold time. The follow-
ing process parameters have been defined in recognised standards:
►
According to WHO: 180 °C with a minimum hold time of 30 min
►
According to the European Pharmacopoeia: 160 °C with a minimum hold time of 120 min
For the inactivation of endotoxin (pyrogenes), dry heat of at least 180 °C can be applied. For
the depletion of pyrogenic substances, you have to keep a combination of temperature and
time going beyond the requirements of sterilisation.
Inactivation of endotoxin is possible using the following process parameters (data in accord-
ance with ISO 20857:2010):
►
180 °C with a minimum effective time of 180 min
►
250 °C with a minimum effective time of 30 min
Caution:
The temperature and time requirements normal for hot air sterilisation do not destroy
endotoxins.
Especially when the appliance is heavily loaded, using these parameters without checking
them will not be sufficient. For safe sterilisation, validation of the individual sterilisation pro-
cess is required. The requirements for the validation of sterilisation by dry heat are e.g. defined
in standard ISO 20857:2010. Also valuable is the "guideline on validation and routine moni-
toring of sterilisation processes using dry heat for medical products" issued by the German
Society for Hospital Hygiene (DGKH).
Sterilisers SF
PLUS
/SN
PLUS
PLUS
PLUS
/IF
PLUS
PLUS
PLUS
/IN
PLUS
PLUS
PLUS
feature
default sterilisation programmes saved in the appli-
ance, which cannot be changed or deleted. A descrip-
tion of how to activate them is provided from page
50, their use is described from page 29. We
recommend to always use these programmes for
sterilisation.
If the process parameters are set manually,
setpoint-dependent operation is automatically
selected.
PROGRAM
Activate
Delete
Test 012
Test 022
Test 013
Test 014
Test 015
Steri 180 °C
Summary of Contents for IF PLUS
Page 59: ......