11
Technical data
Appliance size
30
55
75
110
160
260
450
750
1060
Appliance width D
1
[mm]
585
585
585
745
745
824
1224
1224
1224
Appliance height E
1
[mm]
707
787
947
867
1107
1186
1247
1726
1726
Appliance depth G
1
(footprint) [mm]
434
514
514
584
584
684
784
784
1035
Depth of door lock [mm]
56
Appliance depth F
1
(including door handle) [mm]
490
570
570
640
640
740
840
840
1091
Chamber width A
1
[mm]
400
400
400
560
560
640
1040
1040
1040
Chamber height B
1
[mm]
320
400
560
480
720
800
720
1200
1200
Chamber depth C
1
[mm]
250
330
330
400
400
500
600
600
850
Chamber volume [litres]
32
53
74
108
161
256
449
749
1060
W
eight [kg]
48
57
66
78
96
110
170
217
252
Power [W]
I...
115 V
, 50/60 Hz
800
900
900
900
900
900
1500
1800
–
230 V
, 50/60 Hz
1600
1000
1400
1400
1600
1700
1700
2400
–
U...
230 V
, 50/60 Hz
1600
2000
2500
2800
3200
3400
–
–
–
115 V
, 50/60 Hz
1600
1700
1800
1800
1800
1800
–
–
–
400 V
, 50/60 Hz
–
5800
2
7000
2
7000
2
3 x 208 V
, 50/60 Hz
–
4800
5700
5700
Current consumption
[A]
I...
230 V
, 50/60 Hz
7.0
4.3
6.1
6.1
7.0
7.4
7.4
10.4
–
115 V
, 50/60 Hz
7.0
7.8
7.8
7.8
7.8
7.8
13.0
15.5
–
U...
230 V
, 50/60 Hz
7.0
8.7
10.9
12.2
13.9
14.8
–
–
–
115 V
, 50/60 Hz
13.9
14.8
15.5
15.5
15.5
15.5
–
–
–
400 V
, 50/60 Hz
–
3 x 8,4
2
3 x 10,2
2
3 x 10,2
2
3 x 208 V
, 50/60 Hz
–
3 x 13.3
3 x 15.1
3 x 15.1
max. number of sliding grids
3
4
6
5
8
9
8
14
14
max. load per sliding grid [kg]
20
30
60
max. load per appliance [kg]
60
80
120
175
210
300
Setting temperature range
I...
+20 to +80 °C
3
–
U...
+20 to +300 °C
3 4
Adjustment precision
I...
0.1 K
–
U...
up to 99.9 °C: 0.1 K above 100 °C: 0.5 K
en