42
D33142 | Date 03/2017
Malfunctions, warning and error messages
Error description
Cause of errors
Rectifying errors
See
Error message E-6 in
the humidity display
HUMIDITY
Set 50
.0%rh
%rh
E-6
Humidity sensor defective
►
No humidity con-
trol possible
►
Contact customer
service
page 2
When switching on
the appliance, the start
animation is displayed
in another colour than
white
►
Cyan
:
Not enough storage
space on the SD card
►
Red
:
The system files could
not be loaded
►
Orange
: The
fonts and images could
not be loaded
Contact customer
service
Contact customer
service
Download the firm-
ware update at mem-
mert.com and install it
page 2
page 2
6.3
Power failure
In case of a power failure, the appliance operates as follows:
In manual mode
After power supply has been restored, operation is continued with the parameters set. The
time and duration of the power failure are documented in the log memory.
In timer or programme mode
In case of an interruption of the power supply of less than 60 minutes, the current pro-
gramme is continued from the point at which it was interrupted. For longer interruptions of
the power supply, all appliance functions (heating, fan etc.) are switched off.
In remote control mode:
The previous values are restored. If a programme has been initiated via remote control, it is
continued.