D32802 | Date 11/2018  

13

Operation and control

4.2  Adjustment options

You can set the following parameters after pressing the corresponding activation key (in 

any sequence):
Temperature

Temperature

Adjustment range: 0 to 40 °C

Heating operation is indicated by the 

 symbol.

Cooling is indicated by the 

 symbol.

°C

22

.4

TEMP

°C

22

.0

Set

Humidity

Humidity
Adjustment range: 10 to 90 % rh

Humidification is indicated by the 

 symbol.

Dehumidification is indicated by the 

 symbol.

FEUCHTE

34

%   

70

 

%rh

Set

rh

LED interior lighting

LED interior lighting
Adjustment range: 0 to 100 % in steps of 1%

LICHT

%

40

Fan speed

Fan speed
Adjustment range: 0 to 100 % in steps of 10%

Make sure that no draught forms in the interior due to the 

settings.

LÜFTER

%

50

Please refer to the operating manual for the constant climate chamber HPP 

for the adjustment options of the timer, the programme operation, tempera-

ture and humidity monitoring, the graphical representation and the menu 

functions.

4.3   Setting compressed air dehumidifaction

The compressed air cannot be dehumidified with the ControlCOCKPIT. Use the pressure 

controller (Fig. 12) instead. For best results when dehumidifying compressed air, set the 

humidity range so that it fluctuates around the setpoint with the smallest amplitude pos-

sible. Use the ControlCOCKPIT (Display "

Graph"

) or the AtmoCONTROL software to track 

the relative humidity in the appliance (Fig. 10 and Fig. 11). 

Fig. 10 

 Incorrect operating pressure (high 

humidity fluctuation)

Fig. 10 

Fig. 10 

 Incorrect operating pressure (high 

 Incorrect operating pressure (high 

    

Fig. 11 

 Correct operating pressure (small 

humidity fluctuation)

Fig. 11 

Fig. 11 

 Correct operating pressure (small 

 Correct operating pressure (small 

Summary of Contents for HPPlife

Page 1: ...www memmert com www atmosafe net 100 ATMOSAFE MADE IN GERMANY SUPPLEMENT TO THE OPERATING MANUAL CONSTANT CLIMATE CHAMBER HPP HPPlife ANIMAL KEEPING CHAMBER HPPlife...

Page 2: ...E mail service memmert com For service enquiries please always specify the appliance number given on the type plate Shipping address for repairs Shipping address for repairs Memmert GmbH Co KG Custome...

Page 3: ...functions described in this manual refer to the latest firmware version Due to individual configurations and sizes illustrations in this manual may be slightly dif ferent from the actual appearance F...

Page 4: ...ption 8 2 2 Working range 2 2 Working range 2 2 Working range 8 2 3 Technical data 9 2 4 Scope of delivery 2 4 Scope of delivery 2 4 Scope of delivery 9 3 Putting into operation 10 3 1 Assembling the...

Page 5: ...2 Product safety and dangers The appliances described in this manual are technically sophisticated manufactured using high quality materials and subject to many hours of testing in the factory They r...

Page 6: ...for working with experimental animals are complied with is responsible for ensuring that unauthorised persons have no access to the appli ance is responsible for ensuring that the maintenance plan is...

Page 7: ...ctions and irregularities The appliance may only be used in a flawless condition If you as the operator notice ir regularities malfunctions or damage immediately take the appliance out of service and...

Page 8: ...more humidity Therefore use the central compressed air dehumidification see page 13 2 2 Working range The temperature humidity diagram Fig 2 specifies at what range of temperature and humidity a perm...

Page 9: ...6 5 115 V 50 60 Hz 13 max number of sliding grids max 3 illuminated max load per sliding grid kg 30 max load per appliance kg 200 Temperature Adjustment range 0 C to 70 C Working range with illuminati...

Page 10: ...it 2 2 Insert the exhaust air unit with the pipe socket into the entry port in the back of the chamber Fig 4 and attach it to the chamber with the eight Torx screws included Fig 5 Make sure that the e...

Page 11: ...ompressed air dehumidification 3 3 Switching on the appliance Switch on the appliance by pressing the main switch on the front of the appliance Fig 8 The appliance displays are in English by default w...

Page 12: ...y adjustment Setpoint and actual temperature display adjustment 3 Fan speed display Setpoint and actual temperature display Fan speed display Setpoint and actual temperature display 4 Activation key f...

Page 13: ...limate chamber HPP for the adjustment options of the timer the programme operation tempera ture and humidity monitoring the graphical representation and the menu functions 4 3 Setting compressed air d...

Page 14: ...ctory is 0 2 bar This is enough to achieve a certified dehumidification of 10 humidity at 10 C under laboratory conditions You can adjust the operating pressure after testing the appliance under the o...

Page 15: ...0 fan speed 20 LED lighting and 22 C for 30 minutes Day simulation 60 fan speed 60 LED lighting and 22 C for 12 hours 5 Night simulation 0 LED lighting and 22 C for 11 hours with unchanged fan speed o...

Page 16: ...emmert Stra e 90 96 D 91186 B chenbach Tel 49 9122 925 0 Fax 49 9122 14585 E Mail sales memmert com facebook com memmert family Die Experten Plattform www atmosafe net Animal keeping chamber HPPlife D...

Reviews:

Related manuals for HPPlife