D32802 | Date 11/2018
11
Putting into operation
3.2 Connecting the appliance
3.2.1 Power and water
Connect the appliance to the power and water supply as described in the operating
manual for the constant climate chamber HPP.
3.2.2 Compressed air
Warning! Only connect compressed air to the gas connec-
tion on the back of the appliance. Introducing other gases
or substances can cause severe damage and considerable
danger (explosion or poisoning).
Use a suitable pressure hose to connect the
chamber to an external compressed air
supply (max. 5 bar) at the rear of the
appliance (Fig. 7).
Only oil-free compressed air may be
used.
See page 13 to see instructions on how
to set compressed air dehumidification .
3.3 Switching on the appliance
Switch on the appliance by pressing the
main switch on the front of the appliance
( Fig. 8 ).
The appliance displays are in English by
default when the appliance is switched
on for the first time. You will find
instructions as to how to change the
language in the operating instructions of
the constant climate chamber HPP.
Hauptschalter
>
Zu der Typenbezeichnung gibt es momentan drei Entwurfsrichtungen,
ich kann Ihnen noch nicht sagen, ob Memmert hierzu schon eine Entscheidung
getroffen hat.
Die hier gezeigt Variante, ist die von uns empfohlene Richtung.
UF 110
PLUS
ON
344
.4
444
4
.
TEMP
Set
°C
°C
FLAP
40
%
TIMER
h
d
10
02
End
Sept.
29
22 24
FAN
%
0
%
100
ALARM
max
444
.4°C
auto
+/-
0.0
K
min
444
.4°C
O
O
ON
N
N
N
O
O
N
O
Manu
44.Sept
Fig. 8
Switching on the appliance
Fig. 7
Compressed air connection (max.
5 bar)
Fig. 7
Fig. 7