D33464 | Date 09/2018
49
Menu mode
CAL 1
+0,5°C
CAL 3
+1,6°C
CAL 2
-0,4°C
Factory calibration
20°C
30°C
40°C
0°C
10°C
Fig. 30
Schematic example of temperature adjustment
Example: Temperature deviation at 30 °C should be corrected.
1. Press the activation key to the right of the
CALIB
display. The display is enlarged
and the temperature adjustment option
is automatically highlighted.
Cal1
°C
-0,2
K
Cal2
100.0
°C
+0,1
K
Cal3
180.0
°C
-0,2
K
JUSTIEREN
Temperatur
letzte Justierung 12.10.2012 12:00
40.0
C
K
0,2
-
C
K
0,1
+
C
K
0,2
-
Calibration
Last updated 12.10.2012 12:00:00
Cal1
Cal2
Cal3
20.0
40.0
5.0
Temperature
Humidity
2. Press the confirmation key repeatedly,
until the calibration temperature Cal2 is
selected.
C
K
0,2
-
C
K
0,1
+
C
K
0,2
-
e
Cal1
Cal2
Cal3
40.0
5.0
20.0
3. With the turn control, set the calibration
temperature Cal2 to 30 °C.
C
K
0,2
-
C
K
0,1
+
C
K
0,2
-
e
Cal1
Cal2
Cal3
40.0
5.0
30.0
4. Save the setting by pressing the confir-
mation key. The corresponding calibra-
tion value is automatically highlighted.
C
K
0,2
-
C
K
C
K
0,2
-
e
Cal1
Cal2
Cal3
40.0
5.0
30.0
0,1
+