D33464 | Date 09/2018
47
Menu mode
7.3.8
Gateway
The setup entry
Gateway
is used to connect two networks
with different protocols.
The gateway is set the same way as the IP address (see
page 43).
7.4
Date and Time
In the
Time
display, you can set date and time, time zone and daylight saving time.
Always set the time zone (and summer time yes/no) before you set the date and time.
Avoid changing the set time after that since this can lead to gaps or overlapping when
recording measured values. If you still need to change the time, you should not run a
programme immediately before or after doing so.
1. Activate the time setting. To do so, press
the activation key on the right side of
the
Time
display. The display is enlarged
and the first adjustment option (
Date
)
automatically highlighted.
Date and time
Time
Time zone
Daylight savings
12
.
05
.
2012
12:00
GMT 01:00
Date
2. Turn the turn control until
Time zone
is
highlighted.
Time
Daylight savings
12.05.2012
12:00
GMT
01:00
Date
Time zone
3. Accept the selection by pressing the
confirmation key.
Time
Daylight savings
12.05.2012
12:00
GMT
Date
Time zone
01:00
4. Set the time zone of the installation site
with the turn control, e.g. 00:00 for Great
Britain, 01:00 for France, Spain or Ger-
many. Accept the selection by pressing
the confirmation key.
Time
Daylight savings
12.05.2012
12:00
GMT
Date
Time zone
00:00
5. With the turn control, select the
Daylight
savings
entry
Time
12.05.2012
12:00
GMT 00:00
Date
Time zone
Daylight savings
Balance
Remote Control Off
Gateway
192.168.5 .1
+30 %