d Over the frequency range 150 kHz to 80 MHz,
field strengths should be less than 3 V/m.
Recommended separation distances between por-
table and mobile RF communications equipment
and the Model PG-IRT1603 Infrared Thermometer
The Model PG-IRT1603 Infrared Thermometer is
intended for use in an electromagnetic environment
in which radiated RF disturbances are controlled.
The customer or the user of the Model PG-IRT1603
Infrared Thermometer can help prevent electromag-
netic interference by maintaining a minimum distance
between portable and mobile RF communications
equipment (transmitters) and the Model PG-IRT1603
Infrared Thermometer as recommended below,
according to the maximum output power of the com-
munications equipment.
For transmitters rated at a maximum output power not
listed above the recommended separation distance
d in metres (m) can be estimated using the equation
applicable to the frequency of the transmitter, where P
is the maximum output power rating of the transmitter
in watts (W) according to the transmitter manufac-
turer.
Rated maxi-
mum output
of transmit-
ter W
Separation distance according to
frequency of transmitter m
150 kHz
to 80
MHz
80 MHz
to 800
MHz
800 MHz
to 2.7
GHz
0.01
0.12
0.12
0.23
0.1
0.38
0.38
0.73
1
1.2
1.2
2,3
10
3.8
3.8
7,3
100
12
12
23
53
UK