Immunitetstest
Power frequency (50/60 Hz) magnetic field IEC
61000-4-8
IEC 60601 test level
30 A/m, 50/60 Hz
Compliance level
30 A/m, 50/60 Hz
Electromagnetic environment-guidance
Power frequency magnetic fields should be at levels
characteristic of a typical location in a typical commer-
cial or hospital environment.
NOTE: UT is the a.c. mains voltage prior to applica-
tion of the test level.
Guidance and manufacturer’s declaration – elec-
tromagnetic immunity
The Model PG-IRT1603 Infrared Thermometer is
intended for use in the electromagnetic environment
specified below. The customer or the user of the
Model PG-IRT1603 Infrared Thermometer should as-
sure that it is used in such an environment.
Immunity test
Conducted RF IEC 61000-4-6
IEC 60601 test level
3 Vrms150 kHz to 80 MHz
6 Vrms 150 kHz to 80 MHz outside ISM bands
Compliance level
N/A
Immunity test
Radiated RF IEC 61000-4-3
IEC 60601 test level
10 V/m 80 MHz to 2,7 GHz