In the log tab, one can control where and if the measured data is stored on the PC. The file can also be opened in
Excel from here.
After measurement of each sub-frame, The data is stored with all measurements of one full frame (768 pixels) in
one line. The next frame is stored on the next line.
Each line starts with a timestamp, followed by the temperature reading of each individual pixel, row by row.
After 768 object temperature readings, follow internal parameters, among which the Ta, which is the sensor
temperature.
The data that is stored is the measurement data AFTER application of the de-interlacing and IIR filter if selected,
but without interpolation.