Appendix C:
OSL Calibration
The SARK-110 provides a calibration procedure to compensate the stray
capacitance and impedance of the external test fixture, e.g. the extension cable.
This should be performed every time the external test fixture is changed and
periodically.
Calibration is performed by using an open circuit, a short circuit, and a calibrated
load with the value of the reference impedance; connected to the external test
fixture. The analyzer performs a series of measurements with these loads,
calculate error coefficients, and these are stored in a file that automatically loads
each time you run the analyzer. These error coefficients are used to correct the
measured values.
The product includes an optional calibration loads kit for the BNC to MCX pigtail
cable adapter to support the calibration process.
The calibration operation is available in the «Setup» «Calibration» «OSL
Calibration».