5.8 Read log files correctly
______________________________________________
The log file contains the following information:
6
Operating pauses
In general, the door should only be leant to during operational pauses in order to reduce wear on the door
seal and to avoid premature failure or sticking.
In the event of longer breaks, such as during vacations, the cooling water supply should be turned off (and
the feed water supply from the water treatment unit if one is connected). We recommend that you do not
disconnect the power supply.
If you allow the autoclave to remain unused for a lengthy period of time, we recommend to perform a
sterilization cycle with an empty sterilization chamber, before normal use again.
---------------------------------------
MELAG Cliniklav 25
---------------------------------------
Program : Universal-Program
134°C wrapped
Date : 30.01.2008
Time : 08:35:50 (Start)
Batch No. : 2
---------------------------------------
AIN6: Conductivity 10 µS/cm
Program step Press. Temp. Time
bar °C min
Start 0.01 53.4 00:00
1.Fractionation
Evacuation -0.92 49.6 01:39
Steam entry 0.40 108.8 06:36
2.Fractionation
Evacuation -0.84 58.1 09:54
Steam entry 0.41 109.8 14:05
3.Fractionation
Evacuation -0.84 57.9 17:47
Steam entry 0.40 109.8 21:48
4.Fractionation
Evacuation -0.84 65.9 25:22
Steam entry 0.41 109.8 29:27
Heat up 2.04 134.4 33:45
Steriliz. begin 2.04 134.4 33:45
Steriliz. end 2.15 135.9 39:15
Press. release 0.19 108.7 40:06
Vacuum-Drying
Drying begin -0.30 99.8 40:31
Drying pressure -0.88 80.9 43:21
Drying pressure -0.91 64.7 46:41
Drying pressure -0.92 55.8 50:01
Drying pressure -0.93 50.5 53:21
Drying pressure -0.93 46.9 56:41
Drying pressure -0.94 44.3 60:01
Drying end -0.82 44.2 60:31
Ventilate -0.29 44.4 61:28
End -0.10 44.5 61:58
---------------------------------------
PROGRAM PROPERLY EXECUTED!
Temperature : 136.1 +0.4 /-0.3 °C
Pressure : 2.17 +0.03/-0.03 bar
Sterilization time: 5 min 30 s
Time : 09:37:49 (End)
---------------------------------------
23 200701063 4.07 4.07
Program
Date
Time of start
Batch number of the day
Conductivity of purified feed water
Program phases with the values for steam
pressure and temperature, and time (relative
to program start)
Final report
Mean sterilization temperature/deviations
Mean sterilization pressure/deviations
Duration of sterilization
Time at end of program
Info-line with total number of batches,
Works number and software version.