User Manual for the Cliniklav
25
5
1
Safety instructions
•
When opening the door, particularly after interrupting the drying process, residual steam can escape
from the autoclave chamber.
•
After opening the door, do not touch any metal surfaces - these will be hot! Danger of burns. Always use
the tray lifter to remove trays, or wear suitable hand protectors when taking out other items.
•
Do not sterilize any liquids with this autoclave. It is not licensed for the sterilization of fluids. Failure to
observe this can result in a delay in boiling, which could result in damage to the autoclave, burns and
other injuries e.g. through splintered glass.
•
Do not sterilize any liquids with this autoclave. It is not approved for the sterilization of liquids. Failure to
observe, these instructions may lead to delayed boiling, damage to the autoclave and burns.
•
The autoclave is intended for use outside the patient environment. The minimum distance to the
treatment area must have a radius at least 1.5 meters.
•
Under current VDE regulations, this autoclave is not suited for use in areas where there are risks of
explosion.
•
In order to ensure effective sterilization with the autoclave observe the instructions in this User Manual,
and in particular ensure that the loading of the autoclave is appropriate for the program selected.
•
The autoclave must only be serviced and repaired by MELAG or by its authorised representatives
(specialist dealers or customer services) using only original parts and following service instructions.
•
Documentation media (computer, CF card reader etc.) must be placed in such a way that they cannot
come into contact with liquids.
•
Before opening the housing always disconnect from the mains power supply!
•
If you intend to install a water treatment unit from another manufacturer, then consult MELAG before you
do so.
•
We recommend the installation of a water leak detector.