- After long hours of work, the sensor can become contaminated and the electronic parameters can deviate. If the reading
value is out of range, you must follow these steps to calibrate it.
1. Purchase standard salt bottles of 33% and 75% humidity.
2.
Please deactivate the auto power off mode and switch on the device as indicated in point 11 of the
previous section.
3.
Remove the protective cap and press the HOLD and ON/OFF buttons simultaneously. "---" will appear
on the LCD screen and "1" will appear on the right side of the screen. Place the device in the standard 33% humidity bottle
for 40 minutes. After 40 minutes, the unit will automatically adjust to the 33% value. Calibration work will end when
performed with two values. After calibration with 33%, perform the same procedure with 75%. If calibration is only
performed with the 33% bottle, turn the unit off manually and then on again to repeat the process once the error message
"Er2" disappears. After the two calibrations, the device will function correctly.
Troubleshooting
Problem
Possible reason
Possible Solution
Display off
No battery.
Check and replace the battery.
Shown on the screen
Low Battery
Replacing the battery
PRECAUTIONS
- Do not drop the product or use it by force.
- Do not disassemble the product to avoid breakdowns.
- Replace the protective cap when not in use.
- Do not place the product with corrosive gas.
- Keep the device clean.
- Avoid dust and water.
- Do not immerse the product in water as it will be damaged.
- Please remove the battery if you are not going to use it for a while
Technical Specifications
Product name
Ambient Moisture Meter
Model
KC-321C
Measuring range Ambient temperature T1
-10ºC~50ºC
Temperature range of the external probe T2
-10ºC~70ºC
Humidity range
0~99%
Accuracy ambient temperature T1
±1ºC
Accuracy of the external probe T2
±1.5ºC
Humidity accuracy
5%~95%±3ºC (25ºC)
Temperature Sensitivity
0.1ºC
Moisture Sensitivity
0.1%
Default temperature unit
ºC
Response time
<80s
LCD Auto Power Off
10s
Auto power off
20 minutes (with auto power off)
Accessories
External temperature probe
Power supply
a batterye 9V (6F22/6LR61)
Intensity of work
<25mA
Operating temperature
0ºC~40ºC
Operating Humidity
5%~95% HR (without condensation)
Storage conditions
-30ºC~60ºC, <85% HR (without battery)
Size of
192mm x 61mm x 31mm
Weight
107g