resistance figures are available and lead length is
measured.
A reading of between 100 m
Ω
and 500 m
Ω
will be
indicated on the broken scale line together with the display
of
. Refer to the appliance manufacturer.
Readings greater than 500 m
Ω
are indicated by t h e
flashing display of the word FAIL.
On the 10 Amp range, the flashing display of
w i l l
appear for resistance greater than 600 m
Ω
, indicating that
the test current is below a satisfactory test level. It is
suggested that the 25 A m p range should then be used.
Insulation Te s t
Class
I
Appliance
1.
Press and hold the Insulation Test switch. The
Insulation screen is displayed.
2.
Insulation resistance value is displayed directly
in M .
3.
Release the Test switch.
As a general rule, Class
I
appliances should have an
insulation resistance value greater than 1 M .
A reading of less than 0.25 M is indicated with the word
FAIL.
Class
II
Appliance
1.
Connect the Earth Bond test lead (or probe if
available) to the 10A/25A terminal.
2.
Press and hold the Insulation Test switch. The
Insulation screen is displayed.
3.
Using the crocodile clip, probe areas of likely
leakage around the insulation areas. i.e. joins /
apertures / metal parts.
4.
Release the Test switch.
5.
Disconnect the Earth Bond test lead.
As a general rule, Class
II
appliances should have an
insulation resistance value greater than 2 M .
A reading of greater than 2 M
is indicated on the solid
scale line together with the word PASS.
A reading of less than 0.25 M is indicated with the word
FAIL.
Appliance Testing
12
Summary of Contents for PAT32 Series
Page 1: ...M PAT32 Series Portable Appliance Testers USER GUIDE GUIDE DE L UTILISATEUR GEBRAUCHSANLEITUNG...
Page 20: ...Notes 20...
Page 21: ...M PAT32 Series Portable Appliance Testers GUIDE DE L UTILISATEUR...
Page 39: ...Notes 39...
Page 40: ...Notes 40...
Page 41: ...M PAT32 Series Portable Appliance Testers GEBRAUCHSANLEITUNG...