4-29
How to proceed
Follow the steps below to determine the depth of a metallic conductor:
Step
Action
1
If the depth of the conductor will be determined with the aid of the
Easyloc
Tx
transmitter, couple the transmitter's signal onto the metallic conductor
in such a way that there is as little signal loss as possible (see section
4.1.2 Options for direct and inductive signal coupling
) and switch on the
transmitter (see section
4.1.3 Initial use of the transmitter
Do not forget to set the transmitter to continuous signal, not a pulsed
signal when measuring depth.
2
Press
to switch on the receiver.
3
Select the desired mode of operation (see section
) by pressing the
button.
The symbol for the currently selected mode of operation is shown in the
lower right-hand corner of the display.
4
With a steady grip, hold the receiver perpendicular to the position of a
known metallic conductor, e.g. above a location which was marked during
the site survey. The signal strength bargraph should show zero
(see section
In doing so, the tip of the receiver should touch the ground.
5
Rotate the receiver (through its own axis) over this location until the
maximum signal strength is indicated (see section
6
Press the
button to begin measuring the depth.
For
automatic
depth measurement
(while working in
TRANSMITTER
or
PROBE
mode)
…
For making a
rough estimate
of
the depth (while working in
RADIO
or
POWER GRID
mode)
…
... the measured depth will be
shown in the display:
The following steps can be skipped.
... the following symbol will appear
in the display:
The measurement has to be
continued from
step 7
.
7
Move slowly to one side until the following symbol appears in the display
and an acoustic signal sounds:
1
3