Megger DLRO 200 User Manual Download Page 5

MAIN MENU SCREEN

This screen provides access to the menuing system, through which you set 
up your instrument and choose the desired test parameters. Navigation of this 
menuing system is by means of the cursor control and Enter key.

At the top of the screen are the three main menu choices; “

TEST

”, 

OPTIONS

” and “

CURRENT

”.

Below this are details of the test current selected (

SET CURRENT

), 

maximum current to be allowed (

CURRENT LIMIT

), the number of stored 

tests and the current date and time.

To perform a measurement is straightforward. 

1.   Use the Left and Right arrows of the two-axis cursor control to  

 

 

highlight the 

TEST

 menu. Use the down arrow to select the  

 

 

required test type. Press Enter.

2.   Use the Left and Right arrows of the two-axis cursor control to  

 

 

highlight the 

CURRENT

 menu. Use the down arrow to select SET  

 

 

and press Enter. Type in the desired current using the numeric keys    

 

on the keypad. When finished press Enter.

 

If the current you have set is greater than the current limit, DLRO200    

 

will beep and clear the 

SET CURRENT

 field. Re-enter the desired  

 

 

current, if necessary increasing the current limit to a value equal to  

 

 

or greater than the desired level. (See separate topic for setting the  

 

 

current limit).

3.   Connect the sample to be tested to the instrument terminals and  

 

 

press the 

TEST

 button. The different test types have slightly  

 

 

different connection requirements, which are detailed, in the  

 

 

sections below.

OPERATION

5

Terminate the supply lead with a plug suitable for your circumstances.

Plug the instrument supply lead into a socket outlet and switch the 
instrument on using the On/Off switch situated on the left side of the 
instrument. The 

POWER

 lamp will illuminate, the instrument will display 

the firmware version number and perform a calibration check. If this is 
successful the display will show “

CALIBRATED

”.  If the check fails the 

display will show “

NOT CALIBRATED

”.   

 

 

After a slight pause the display will change to the Main Menu screen.

DLRO 200       V1.00

COPYRIGHT 

MEGGER LIMITED

 2002

INITIALISING HARDWARE

CALIBRATED

TEST  OPTIONS  CURRENT

SET CURRENT   150 

CURRENT LIMIT 200 

23 TESTS STORED

21/03/02   15:13

200

200

www.actoolsupply.com

www.actoolsupply.com

Summary of Contents for DLRO 200

Page 1: ...M DLRO 200 High Current Low Resistance Ohmmeter USER MANUAL www actoolsupply com www actoolsupply com...

Page 2: ...a voltage upon it greater than 10 volts ac peak or dc with reference to ground the DLRO200 will indicate the presence of the voltage and prevent any tests being carried out n Testing inductive circui...

Page 3: ...tives CONTENTS 3 SAFETY WARNINGS 2 GENERAL DESCRIPTION 4 OPERATION 5 Main Menu Screen 5 TEST Menu 6 OPTIONS Menu 7 CURRENT Menu 9 Warning Messages 9 Entering Notes In The Memo Screen 9 Test Result Sto...

Page 4: ...nt All four terminals are floating A warning will be given if dangerous voltages are connected to any terminal C or P with respect to ground GENERAL DESCRIPTION 4 The Megger DLRO 200 is an a c powered...

Page 5: ...LRO200 will beep and clear the SET CURRENT field Re enter the desired current if necessary increasing the current limit to a value equal to or greater than the desired level See separate topic for set...

Page 6: ...ely connected to the item under test before pressing the Test button The DLRO200 checks voltage and current loop continuity as in Normal Mode and if adequate passes a continuous d c current repeating...

Page 7: ...sing the two axis cursor control and the Enter key These are Retrieve Passbands Set Clock Delete Data and Storage Retrieve allows the recall of results which have been stored within the DLRO200 s inte...

Page 8: ...ng may continue without Passbands The passband option allows you to set upper and lower limits between which the test result must lie if it is to be assigned a Pass Readings that fall above the upper...

Page 9: ...sed Initially the Potential circuit is checked If there is poor continuity in this circuit a message will appear on the display stating POTENTIAL FAIL Press Enter to clear this message and correct the...

Page 10: ...u do not wish to add notes press the Test button and a new test will be started or the Enter button to return to the Main Screen TEST RESULT STORAGE Each test is numbered sequentially dated and timed...

Page 11: ...r these leads should be adequately insulated and sufficiently strong mechanically to support their own weight if long leads are being used The size of the current leads will limit the maximum resistan...

Page 12: ...ent Ranges Full Scale Resolution Volts full scale Max Test Current 999 9 0 1 0 2 V 200 A 9 999 m 1 2 V 200 A 99 99 m 10 4 V 40 A 999 9 m 100 4 V 4 A Subject to supply voltage conditions see below Accu...

Page 13: ...M Megger is a registered trademark DLRO200_UG_en_V04 0810 www actoolsupply com www actoolsupply com...

Reviews: