SAFETY WARNINGS
THESE SAFETY WARNINGS MUST BE READ AND UNDERSTOOD BEFORE THE INSTRUMENT IS USED.
n
DLRO200 is supplied without a plug terminating the supply lead.
An appropriate plug must be fitted prior to use.
n
The DLRO200 must be grounded (earthed) when in use.
If the ground / earth in the socket outlet being used to energise the DLRO200 cannot be relied upon, the instrument must be connected to a reliable
earth / ground via the large green and yellow terminal on the left hand side of the instrument. If no ground / earth is present the instrument will refuse to
switch on and the display will remain blank.
n
Circuits must be de-energised before testing.
The DLRO200 must only be used on dead systems. If the item under test has a voltage upon it greater than 10 volts ac peak or dc with reference to
ground the DLRO200 will indicate the presence of the voltage and prevent any tests being carried out.
n
Testing inductive circuits can be hazardous:
The DLRO200 is a high power instrument, designed for testing resistive loads. It must NOT be used to test inductive loads.
n
During use the DLRO200, its current leads and the sample being tested can become hot.
This is normal and is caused by the passage of high currents. Take care when touching the instrument, current leads, clips and the test sample.
n
This product is not intrinsically safe. Do not use in an explosive atmosphere.
n
Ensure that the power switch is always accessible so that power can be disconnected in an emergency.
n
If this equipment is used in a manner not specified by the manufacturer, the protection provided by the equipment may be impaired.