4.1 ECG Electrode Placement 

 
The eMotion Faros Sensors are compatible with 4 mm snap connector ECG Electrodes, like Ambu 
Blue Sensor M-00-S or VLC-00-S. Place the electrodes on the subject according to the instructions 
illustrated in Figure 12-14 and connect the snap connector electrode wires to the electrodes. 
 
 

The user should be familiar with correct placement of electrodes. Failure to correctly place 
the electrodes will weaken the data acquisition reliability. 
 
 

The quality of an ECG measurement depends on preparation and the skin-electrode 
resistance value. To ensure quality of ECG and to minimize skin-electrode resistance, 
follow these casual instructions. 
 

 

Place electrodes on intact, clean and dry skin. Shave off excess body hair from electrode 

are, if needed 

 

Thoroughly clean the area with alcohol, 

 

Let area dry before applying electrodes. 

 

 

Figures 12-14 presents illustrative position of electrode placement 

 

A and B set-up are for 1-channel ECG data acquisition.  

 

C set-up is for 3-channel ECG data acquisition with Faros 360. 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

NOT

E!

Summary of Contents for eMotion 180 FAROS

Page 1: ...800778 1 4 EMOTION FAROS SERIES MANUAL DOC DATE OF ISSUE DECEMBER 15 2015 Mega Electronics Ltd 2014 2015 800778 1 4...

Page 2: ...echnical Specification 9 2 4 Sensor Light Indicators and States 10 2 5 Recharging the Battery 11 2 6 Troubleshooting 13 3 Measurement Preparation 14 3 1 Offline ECG Data Acquisition 15 3 2 Online ECG...

Page 3: ...from the sensor 35 How can I reset my eMotion Faros device 35 How can I change the measurement modes of the device 35 Which data format I should choose for my measurement SDF Suunto Data Format or EDF...

Page 4: ...data acquisition or communicates remotely with Bluetooth enable device ECG and HRV measurements can be done by using Faros Sensor with Stingray adapter and Textile Belt or Cable Set Fast Fix Electrode...

Page 5: ...diagnosis or self treatment based on acquisitioned data Not suitable to use in MRI environment The device is not intended to be used at the same time with high frequency HF surgical equipment or with...

Page 6: ...ts means that additional instructions that further explain use of a particular part or function is found in the Device Manual or Service Manual The Lot number follows this symbol Electrodes General Wa...

Page 7: ...ivity monitoring Sensor Specifications Simple stress and Recovery Monitoring Heart rate variability Physical Activity ECG offline holter 1 GB memory Memory capacity up to 2 months measuring Rechargeab...

Page 8: ...and Textile Belt are used for one channel ECG acquisition For further information see chapter 4 1 Textile Belt 900810 Stingray Adapter 900809 Electrode Leads are connected to electrodes placed on pati...

Page 9: ...SDF ASC and EDF SDF ASC and EDF SDF ASC and EDF Power source 3 7 V Li ion battery 3 7 V Li ion battery 3 7 V Li ion battery Dimension 48 x 29 x 12 mm Weight 13g 48 x 29 x 12 mm Weight 13g 48 x 29 x 1...

Page 10: ...ontinuously Measurement start Green light until the signal is recognized Measurement Measurement and Power on Indicator light blinks Measurement battery near empty Caution Indicator is blinking at the...

Page 11: ...llowed in Sensor s USB connector area Check the Sensor s USB connector for dust and remove it with scrap paper towel If there is moisture on sensor let the sensor dry 2 hours on table before you plug...

Page 12: ...s start to blink Please note that Charger Station s battery charges about 15 minutes before sensor recharging starts if station s battery is completely empty Charger station can also be used to recha...

Page 13: ...ent error light indicator blinks three 3 times Recharge device Synchronize device clock via Faros Device Manager Save settings Reset the device Push the pushbutton until all light indicators flash Pow...

Page 14: ...ros Sensor to computer via USB Cable and wait until Windows has finished the installation of the device drivers Check that Measurement and Power on Indicator shine continuously Browse to FAROS_xxx har...

Page 15: ...logger mode from Manager view 3 Select wanted acquisition parameters RR Intervals Accelerometer ECG 4 Select values for acquisition parameter Accelerometer Sampling Accelerometer Dynamic Range ECG Sam...

Page 16: ...C Virhe M rit Aloitus v lilehdess Heading 1 P otsikko jota haluat k ytt t h n kirjoitettavaan tekstiin 16 40 Figure 7 Example Configuration for ECG data logging with Faros 180 Figure 8 Example Configu...

Page 17: ...Manager eMotion Faros Manager is locate on Faros internal memory 2 Select Online Mode from Manager view 3 Press Save to apply configuration Now your Faros sensor is ready to perform Online ECG data mo...

Page 18: ...Faros internal memory 2 Select Datalogger mode from Manager view 3 Set wanted acquisition parameters RR Intervals Accelerometer ECG 4 Select values for acquisition parameter Accelerometer Sampling Acc...

Page 19: ...to place electrodes on intact clean and dry skin Shave off excess body hair if needed Place Cable Set or Stingray Adapter to Faros Sensor connector Place Cable set or Stingray Textile Belt on subject...

Page 20: ...th correct placement of electrodes Failure to correctly place the electrodes will weaken the data acquisition reliability The quality of an ECG measurement depends on preparation and the skin electrod...

Page 21: ...MANUAL DOC 21 40 Virhe M rit Aloitus v lilehdess Heading 1 P otsikko j ECG Electrode Placements Figure 12 2 electrode cable set placement Placement R RA IEC Colour Red AHA Colour White Placement F LL...

Page 22: ...ehdess Heading 1 P otsikko jota haluat k ytt t h n kirjoitettavaan tekstiin 22 40 Figure 13 3 electrode cable set placement Placement R RA IEC Colour Red AHA Colour White Placement F LL IEC Colour Gre...

Page 23: ...v lilehdess Heading 1 P otsikko j Figure 14 5 electrode cable set placement Placement R RA IEC Colour Red AHA Colour White Placement F LL IEC Colour Green AHA Colour Red Placement L LA IEC Colour Yel...

Page 24: ...Textile Belt Moisten well the belt s electrode areas with water Wear textile belt under chest muscle line as shown in Figure 17 Check that the wet electrode areas are against your skin and that the S...

Page 25: ...800778 1 4 EMOTION FAROS SERIES MANUAL DOC 25 40 Virhe M rit Aloitus v lilehdess Heading 1 P otsikko j Figure 17 Correct Placement of Stingray and Textile Belt...

Page 26: ...ect Sensor to USB connector Remove liners from contact tapes and place sensor on electrode Figure 18 Connecting Sensor to electrode Peel the adhesive area liner away from the electrode Do not touch th...

Page 27: ...g vibration of a car The total acceleration is represented in the three accelerometer channels x y and z as vector components The direction of each acceleration component is represented as the sign of...

Page 28: ...Browse to FAROS_xxx hard drive 3 Open Software Folder 4 Double click eMotion EDF Viewer folder and run eMotionEDFViewer application It s recommended to install eMotion EDF Viewer to computer before us...

Page 29: ...elect desired measurement file a Browse FAROS_xxx hard drive and open DATA folder Folder names indicate the date of performed measurement s b Browse file location from your personal computer 3 Select...

Page 30: ...ta recorded by the accelerometer is used to recognize the movement periods e g sports walking or other physical activity from the measurement The accelerometer data itself does not have an effect on t...

Page 31: ...dess Heading 1 P otsikko j User Interface Data window presents the acquisitioned ECG data Possible events are shown in data Info window contains Faros sensor configuration which was used in measuremen...

Page 32: ...vents ECG event report is created of inserted events i e markers Click Event Report to generate repor of desired events t A PDF report will be open in separate window having options to print or save 5...

Page 33: ...us v lilehdess Heading 1 P otsikko j Calibrate To calibrate Data window mm grid view select Calibrate from Tools Measure the blue rectangle and write the values below Click OK Figure 26 Calibration vi...

Page 34: ...ros ECG Sensor and Stingray Adapter The eMotion Faros ECG Sensor and Stingray adapter can be cleaned by wiping device with using a non fluffing cloth dampened in cleaning fluid such as mild hand soap...

Page 35: ...nnect the device with PC via USB you can browse to the hard drive of the device named FAROS_ and copy cut and paste the needed files to your computer If you wish you can also open the measurement file...

Page 36: ...o reset the device If resetting the device does not help please contact mega megaemg com for support What is the suitable sampling frequency when measuring ECG This depends significantly on the purpos...

Page 37: ...ipment Part 1 General requirements for safety EN60601 1 2 3rd edition IEC601 1 2 3rd ed Medical electrical equipment General requirements for safety 2 Collateral Standard Electromagnetic compatibility...

Page 38: ...itoring biofeedback and scientific research The eMotion Faros ECG Mobile is intended for use in clinical and non clinical settings to collect and transmit health parameters to healthcare professionals...

Page 39: ...800778 1 4 EMOTION FAROS SERIES MANUAL DOC 39 40 Virhe M rit Aloitus v lilehdess Heading 1 P otsikko j BLANK PAGE...

Page 40: ...v lilehdess Heading 1 P otsikko jota haluat k ytt t h n kirjoitettavaan tekstiin 40 40 Manufacturer Mega Electronics Ltd Pioneerinkatu 6 70800 Kuopio Finland Phone 358 0 17 581 7700 Fax 358 0 17 580...

Reviews: