Electromagnetic Emissions
Emissions Test
Compliance
RF emissions
EN 60601-1-2:2007+AC:2010, IEC 60601-1-2:2007, CISPR
11:2009+A1:2010 and IEC 60601-1-2:2014
Group 1
RF emissions
EN 60601-1-2:2007+AC:2010, IEC 60601-1-2:2007, CISPR
11:2009+A1:2010 and IEC 60601-1-2:2014
Class B
Electromagnetic Immunity
Immunity Test
IEC 60601 Test Level
Compliance Level
Electrostatic
discharge
(ESD)
IEC 61000-4-2
±2.0 kV, ±4.0 kV, ±6.0kV,
±8.0 kV contact discharge
±2.0 kV, ±4.0 kV, ±8.0 kV,
±15.0 kV air discharge
±2.0 kV, ±4.0 kV, ±6.0
kV, ±8.0 kV contact
(56% RH)
±2.0 kV, ±4.0 kV, ±8.0
kV, ±15.0 kV air (56%
RH)
RF electromagnetic field
immunity test
IEC 61000-4-3
10 V/m
10 V/m
Power frequency magnetic
fields
IEC 61000-4-8
30 A/m
30 A/m
Warning
:
1. The TouchCare® transmitter should be avoided to use in an environment with high
voltage, high-intensity magnetic field, where the intensity of EM DISTURBANCES is
high.
2. Portable RF Communications equipment should be used no closer than 30 cm (12
inches) to the TouchCare® transmitter. Otherwise, degradation of the performance
of the transmitter could result.
3. Use of this equipment adjacent to or stacked with other equipment should be
avoided, because it could result in improper operation. If such use is necessary, this
equipment and other equipments should be observed to verify that they are
operating normally.