Do NOT use household cleaners, chemicals, solvents, bleach, scouring pads or sharp 
instruments to clean your transmitter. Small amounts of rubbing alcohol can be used 
to  clean  the  surface  of  your  transmitter.  Never  put  your  transmitter  in  the 
dishwasher or use very hot water to clean it. 

Do  NOT  use  a  hair  dryer,  microwave  oven,  or  conventional  oven  to  dry  your 
transmitter. Use a soft towel. 

Do NOT clean your transmitter while it is in use.   

Please  remove  the  sensor  if  redness,  bleeding,  pain,  tenderness,  irritation,  or 
inflammation develops at insertion site, of if you experience unexplained fever. 

The  sensor  may  create  special  needs  regarding  your  medical  conditions  or 
medications. Discuss with your healthcare providers before using the sensor. 

If  failure  or  damage  of  your  transmitter  is  found  during  usage,  please  contact 
customer support or your local Medtrum distributor for replacement. 

Radio Frequency (RF) Communication 

Note:  The  TouchCare®  transmitter  can  generate,  use,  and  radiate  radio  frequency 
energy, and may cause harmful interference to radio communications. There are no 
guarantees  that  interference  will  not  occur  in  a  particular  installation.  If  the 
TouchCare®  transmitter  does  cause  harmful  interference  to  radio  or  television 
reception,  you  are  encouraged  to  try  to  correct  the  interference  by  one  of  the 
following measures: 

• Move or relocate the TouchC

are® transmitter. 

• Increase the distance between the TouchCare® transmitter and the other device 

that is emitting/receiving interference. 

Common  consumer  electronic  devices  that  transmit  in  the  same  frequency  band 
used  by  the  TouchCare®  transmitter  may  prevent  communication  between  the 
transmitter  and  the  receiver.  This  interference,  however,  does  not  cause  any 
incorrect data to be sent and does not cause any harm to your device. 

RF communication between your transmitter and receiver works up to a distance of 
10 meters (33 feet). 

Charge the Transmitter 

The  TouchCare®  transmitter  is  charged  via  a  USB  charging  cable  which  is  plugged 
into a USB 2.0/3.0 port or a power adapter with a rated voltage of DC 5 V and a rated 
current higher than DC 100 mA. The device with the USB port and the power adapter 
must comply with EN 60950-1 or EN 60601-1. 

The  battery  must  be  fully  charged  the  first  time  using  the  transmitter,  which  may 
take up to 2 hours. It is recommended to recharge the transmitter after each sensor 