Fill the New Reservoir Patch
With the help of your healthcare provider, decide on the amount of insulin (70 U - 200 U) you need to insert
into the Reservoir Patch.
5.Slowly fill the Reservoir
with insulin
2. Fill the syringe with insulin.
1. Push air into the vial.
Remove air
3. Keep the syringe vertical
to the Patch Pump, and
the needle inside the fill
port. Retract the plunger
to remove air bubbles.
4. Push the plunger
until you see a drop
of insulin at the tip of
the needle
Prime the Pump
6
Press button cover.
Leave the button cover on until priming is completed.
Warning:
Do not inject air into the fill port. Doing so may result in unintended or interrupted insulin delivery.