background image

LSUHSC Shreveport 
Radiology Department 
Proc 13.17 

Condition #7 

When the 
injection starts, 
almost no 
pressure buildup 
is witnessed, 
when a buildup 
was expected. 

Due to operator error, there 
may be no fluid in the 
syringe (you could be 
injecting air).  The injector 
will experience almost no 
resistance to moving the 
syringe plunger forward 
with an empty syringe. 

 

Note

:  Very slow injections 

through large disposable 
sets will require and display 
very little pressure.  This is 
normal.  What is being 
described above is a 
condition where a buildup 
of pressure was expected, 
and none was witnessed.  
OR:  You may have two 
syringes on the injector, 
only performing a single 
syringe injection, and the 
LPCT may be connected to 
the wrong syringe.   

Stop the injector 
immediately and 
investigate. 
 
 
Check the fluid path for 
air.  
 

Condition #8 

At the end of a 
normal injection, 
the pressure 
graph does not 
immediately 
drop to zero. 

Stellant is designed to inject 
all of the contrast in the 
syringe to help minimize 
contrast waste.  Sometimes 
in the process of doing this, 
the piston pushes against 
the front of the syringe as it 
squeezes out the last drops 
of fluid.  This in turn 
generates pressure, which is 
displayed by the Pressure 
Graph.  This pressure 
reading should drop to 
baseline after about 50 
seconds.  

No action. This is normal 
operation.   

NOTE:  Medrad makes no claims about the Pressure Monitor Graph other than it is a 
graphic representation of the force required to move the syringe plunger, which closely 
approximates the pressure developed in the syringe.  
 
 
 

Written: 1/19/2007 

Summary of Contents for Stellant D

Page 1: ...he hand on the ace of the gantry during injection This allows for uninterrupted passage of injected contrast through the axillary and subclavian veins at the thoracic outlet 7 A small volume test inje...

Page 2: ...wly can result in an error message Piston will automatically advance to the front of the syringe SETUP 1 Using sterile technique attach the sterile loading device and load the syringes by pressing AUT...

Page 3: ...the remaining volume 4 Armed Indication Lights Contrast green Saline blue will flash when the injector is armed or on hold and will be solid when injecting 5 Initiate the injection at the patient sid...

Page 4: ...essage cannot be corrected and or the system is not Operating properly DO NOT USE the system Call MEDRAD Service c Routine inspections and maintenance will ensure continued performance and reduce equi...

Page 5: ...theter is being used An additional disposable is in the fluid path causing a greater restriction to the flow of fluid extension tube stopcock check valve IV extension set Defective syringe under lubri...

Page 6: ...d causing this condition See Pressure Limiting in the Operation Manual for a thorough discussion of causes and resolution Note Medrad makes no claims about the information on the Pressure Monitor Grap...

Page 7: ...onnector Tubing extension set or catheter has burst or disconnected from the fluid path fluid is running on the floor The syringe may have failed Fluid is being forced around the syringe plunger inste...

Page 8: ...nd the LPCT may be connected to the wrong syringe Stop the injector immediately and investigate Check the fluid path for air Condition 8 At the end of a normal injection the pressure graph does not im...

Reviews: